Section 3.  Scale Operation 

 

 

 

Fig. 3 Front Panel 

 
 
 

Display Pushbuttons 

The Model DS6150 controls consist of UNITS, ZERO, REWEIGH, 

▲, ▼, and 

BMI buttons.  The display is used to provide weight indications and operator 
messages indicating scale operation.  

 

 

Power On/Off 

Press the ZERO button to turn on the indicator. Press and hold the ZERO button 
for 3 seconds to turn off the indicator. The indicator has an adjustable Automatic 
Shutoff Timer that will turn off the indicator when not in use. 

 

Zero 

When the ZERO button is pressed, the scale will reset to zero weight.  This 
feature can be used to zero out any weight not desired to be part of the patient 
weight. The current BMI entry will be changed back to a 0.0 value.  

 
Units 

Press the UNITS button to toggle between lb and kg weighing units.  The current 
unit will be displayed to the right of the weight digits.  

NOTE: The UNITS button 

Содержание DS6150

Страница 1: ...Model DS6150 Remote Indicator Scale Technical Manual Doran Scales Inc 1315 Paramount Pkwy Batavia IL 60510 USA 1 800 264 4107 www doranmedical com MAN0315 Revision 0 1 2 12 2014...

Страница 2: ......

Страница 3: ...5 Parameter Setup 12 Entering and Exiting Setup Mode 12 Calibration Mode Entry 12 Reset All Parameters to Default Settings 13 Automatic Shutoff Timer 13 Average Aperture 14 Data Output 14 EMR Format...

Страница 4: ...y not be covered by the warranty Please observe the following precautions to insure years of trouble free service from your new scale Do not use sharp objects to press any of the buttons Do not pick u...

Страница 5: ...2 Wall Mounting If mounting the indicator to a wall refer to the following diagram Place two screws 4 09 apart 65 above the floor The screw heads should be approximately 0 33 in diameter...

Страница 6: ...to release battery compartment cover Completely remove all the batteries from the indicator and install the new set of batteries Be sure the batteries are oriented properly Reinstall the battery comp...

Страница 7: ...4...

Страница 8: ...r 3 seconds to turn off the indicator The indicator has an adjustable Automatic Shutoff Timer that will turn off the indicator when not in use Zero When the ZERO button is pressed the scale will reset...

Страница 9: ...e Hold Indicates a stable weight has been achieved and held on the display This indicates an accurate weight has been achieved Note Weights below 4 5 pounds will not be held Average Hold Indicates tha...

Страница 10: ...mber and Revision display Pressing and holding the REWEIGH button for 3 seconds will perform a display test The software number Su179 and the revision r 3 2 or higher are displayed first After display...

Страница 11: ...econds BMI Operation 1 With the held weight displayed press the BMI button to enter the height value Height entry starts at 5 6 0 or 168 0 representing 5 foot 6 inches or 168 centimeters NOTE Height u...

Страница 12: ...or by using your own body weight Two Point Calibration Procedure 1 When the setup mode is entered the display reads CAL n 2 Press the button to change the display to CAL y 3 Press BMI 4 The display r...

Страница 13: ...e sure vibration and air currents are not present 9 Press BMI to perform a zero point calibration The display will count down to zero The counter will reset if too much motion Er nn is detected 10 The...

Страница 14: ...alibration has detected a negative range If this error persists consult the raw counts table The following table shows the acceptable Raw Counts for no load and full load Raw Counts can be viewed thro...

Страница 15: ...gation 1 Press BMI to scroll forward to the next parameter 2 Press or buttons to change the current parameter setting 3 Press Reweigh to save any changes and exit setup mode Calibration Mode Entry Cal...

Страница 16: ...ration is performed Reset All Parameters to Default Settings rSt Default parameters n No Parameters are not defaulted y Yes Set all parameters to the factory default values when BMI is pressed Calibra...

Страница 17: ...n a request command is received through the RS 232 or USB port A 5 Transmits an EMR string after the STABLE HOLD or AVERAGE HOLD weight is displayed default A b Transmits an EMR string after the Heigh...

Страница 18: ...disabled default LB OZ Units Enable Disable lb oz units Enable Disable on lb oz weighing units active of lb oz weighing units disabled default Inches Units Enable Disable Inches Height units Enable D...

Страница 19: ...disabled y Beep enabled default Battery Voltage Level DC Displays current battery voltage DC Battery Bar Indications Empty Battery Symbol 5 1 5 9 V One Bar 6 0 6 9 V Two Bars 7 0 7 9 V Three Bars 8 0...

Страница 20: ...ation of that system These are Baud Rate Data Bits and Parity The baud rate determines how fast the data is sent from the scale The sending and receiving units must be set to the same baud settings Re...

Страница 21: ...ng the Reweigh button Welch Allyn Spot protocol VU 2400 7 EVEN 1 Command RXD Computer to Scale Scale Output Responds TXD Scale to Computer Description DR cr lf mot sp sp weight uu cr lf Current Weight...

Страница 22: ...will wait until stable then zero cr or W bmi height pol gross uu sp G BMI uu sp T pol net uu sp N cr lf Current Display Weight Active Stable Hold or Average Hold mot G or N stable g or n motion U Scal...

Страница 23: ...ble character cr ASCII 0dh Carriage return non printable character lf ASCII 0ah Linefeed non printable character sp ASCII 20h space mot GR for stable gr for motion or SD for stable S_ for motion pol m...

Страница 24: ...6 5 W x 2 3 H Power 12VDC AC Adapter Negative Center Optional 6 AA Alkaline Batteries Displayed Units lb kg Capacity 500 x 0 2 lb or 230 x 0 1 kg Construction Painted Mild Steel Frame Stainless Steel...