Calibration Troubleshooting
If
gruL
or
groL
appears on the display, the calibration zero is out of range.
Check connections between the indicator and base. Press BMI to perform the
zero calibration again. If this error persists, consult the raw counts table.
If
ErSp
appears on the display, the span calibration is out of range. Place the
proper weight on the platform and press BMI to perform the span calibration
again. If this error persists, consult the raw counts table.
If
Erng
appears on the display, the span calibration has detected a negative
range. If this error persists, consult the raw counts table.
The following table shows the acceptable Raw Counts for no load and full load.
Raw Counts can be viewed through the last setup parameter. Refer to Section
5, Parameter Setup, for more information regarding parameter viewing.
If the scale’s raw counts are not within the ranges specified in the following table,
contact Doran’s Technical Assistance department at 800-264-4107.
Calibration Requirements in Raw Counts
Minimum
Maximum
Zero Calibration Point
-2000
20000
Required Calibration Span
(100% of Capacity Scale Cal Point
– Zero Cal Point)
6000
30000
Required Calibration Span
(5% of Capacity Cal Point
– Zero Cal Point)
300
1500