3.4 Service Procedures
It is essential that the generator is isolated from the mains power supply and fully
depressurised before carrying out any servicing procedures.
Stopping the equipment
Close the ball valve on the N2 Outlet port.
Press [ ] to initiate the shutting down sequence.
The generator will complete the current cycle and then exhaust both beds.
Observe the column pressure gauges on the side on the generator to ensure that they read 0bar.
When the generator is completely depressurised it will revert to standby mode.
Disconnect the generator from the electrical supply at the supply disconnection switch. Ensure that this switch is locked off to prevent
accidental operation.
Disconnect the compressed air supply from the generator.