Prior to shipment the 750 TOBS has gone through 2 days of continuous testing, including a final
force transducer calibration. However, to ensure that the Tissue Organ Bath is working at its highest
performance, DMT recommends that a new force transducer calibration is performed before the first
use of the 750 TOBS system.
As part of the general maintenance of the 750 TOBS system, DMT also recommends that the bath is
weight calibrated at least once every month, every time the system has been moved or has not been
used over a long period.
The weight calibration procedure is described in detail in chapter 4 in the User Manual.
2.4 ISOMETRIC TRANSDUCER:
Figure 4.1.1: Transducer with calibration weight.