CHAPTER 2
Tissue holding pins from top view
Tissue holding pins from side view
Figure 2.3
- Illustrations of properly aligned tissue holding pins (depicted on the far left) and
incorrectly aligned pins (depicted in the middle and far right).
Figure 2.2
- Illustration of the screws for fine adjustment of the pins
A B E D C
2.1.2 Fine adjusting the pins for larger vessels (figure 2.2 and 2.3)
1.
Loosen screw “A” to move the micrometer-side arm holder sideways
2.
Loosen screw “B” to move the micrometer-side pin toward or away from the transducer.
3. Loosen screw “C” to align the transducer-side tissue holding pin horizontally.
4.
Loosen screws “D” and “E” to align the heights of the tissue holding pins vertically.
2.2 Calibration of the force transducer
As a part of the general maintenance of the Tissue Bath, DMT recommends that the Tissue Bath is force calibrated at least once a
month. The Tissue Bath should also be force calibrated every time the Multi Interface has been moved. Although lab benches are
all supposedly perfectly horizontal, small differences in lab bench pitch can affect the calibration of the system. The Tissue Bath
should also be calibrated if the system has been idle for longer than a month. A step-by-step procedure is explained in chapter
3.5.1.2 in
Multi Myograph System - User Manual
.