16
4.2 Force Transducer Replacement
If the force transducer breaks and needs to be replaced the 560TP need to be shipped back to DMT for replacement. The ship-
ping address is:
Danish Myo Technology A/S
Skejby Science Center
Skejbyparken 152
DK-8200 Aarhus N.
Denmark
Tel: +45 8741 1100
Please contact your Sales Representative and request a Return to Manufacturer Authorisation (RMA) Number to attach to your
shipping documentation.
4.3 Maintenance
The 560TP is a very delicate and sophisticated piece of research equipment. DMT recommends that the following sections are
read carefully and that the instructions are followed at all times.
Force transducer:
The force transducer is the most delicate and fragile component of the 560TP. Extreme care must be used
when handling or touching the force transducer. As part of daily maintenance, inspect the grease around the transducer pin ex-
tending from the transducer compartment pinhole before starting any experiment. Insufficient grease in this area will allow buffer
and water to enter the transducer compartment and cause irreparable damage to the force transducer.
IMPORTANT NOTES
• DMT RECOMMENDS THAT THE HIGH VACUUM GREASE SEALING THE TRANSDUCER PINHOLE IS CHECKED AND RE-
PLACED AT LEAST ONCE A WEEK, ESPECIALLY IF THE MYOGRAPH IS USED FREQUENTLY
• DMT TAKES NO RESPONSIBILITY FOR THE USE OF ANY OTHER KIND OF GREASE OTHER THAN THE GREASE SUPPLIED
BY DMT FOR THIS PURPOSE
• DMT TAKES NO RESPONSIBILITY FOR ANY KIND OF DAMAGE CAUSED BY MISHANDLING OF THE FORCE TRANSDUCERS