10
Figure 2.4
Puller Controller close-up
CHAPTER 2
Figure 2.5
‘Current position’ box
2.4 Force Calibration
Before the 560TP is packed and shipped it has gone through two days of continuous testing, including a final force transducer
calibration. However, DMT recommends that a force transducer calibration is performed before using the 560TP System for the
first time. The force transducer calibration is described in the following section.
1. Click on the ‘Calibrate Force’ button on the right hand side of the MyoPULL software window (see figure 2.5).
2. Read the calibrate information on the screen carefully. Please note that even very small vibrations or slight instability of
mounting supports will affect the calibration.
3. Zero the force by clicking the
‘
Zero Force
’
button (see figure 2.6). The force transducer is now zeroed.
4. Press ‘Next’.
Figure 2.6
‘Zero Force’ button and ‘Next’ button