29
7.3 Monthly
-
Use the non-corrosive cleaner and stiff bristled brush or sponge to clean the water level
sensor at the rear of the chamber as Figure 30.
Caution:
Clean the dirt off from the sides of the sensor is more important than the tip. Use a
damp cloth to wipe the surface after cleaning.
Figure 30
-
Clean the chamber and piping system with “CHAM-MATE” following the instructions on the
sachet.
-
Check the safety valve
Turn off the power and unplug the sterilizer. Remove the water reservoir cap as Figure 31.
Use a screw driver to pull the metal ring of the safety valve for approx. 3 seconds; then
release. Perform the check 3 times. Put the water reservoir cover back.
Warning:
If excessive force is required to pull the safety valve, it must be replaced. Call for
service.
Figure 31
Metal Ring
Water Level Sensor