1. Important Safety Instructions
Caution:
Please install, operate and maintain the sterilizer in accordance with this
Instruction Manual. Failure to do so could result in serious injury or damage to the
unit.
Warning:
The outer casing and metal surfaces of the sterilizer will be hot during operation,
please do not touch it.
Warning:
Steam and hot water will be present when opening the door after a sterilizer cycle.
Avoid contact.
Warning:
DO NOT place alcohol or other flammable items in the sterilizer. An explosion
could occur, causing personal injury.
Warning:
DO NOT place any objects on the top of the sterilizer.
Warning:
It will require at least two (2) or more people to carry the sterilizer to avoid
dropping it off by mistake.
Warning:
Always check the pressure gauge before opening the door. DO NOT attempt to
open the door if the pressure is not at zero (0).
Warning:
Use only distilled water. Normal tap water contains minerals, especially chlorides,
which have corrosive effects on stainless steel. Failure to use distilled water will
invalidate the warranty.
Warning:
Always allow a minimum of 20 minutes between each sterilization cycle.
Warning:
The door must be closed completely during operation of the unit. If the “DOOR”
indicator light illuminates, it means that the door is not closed properly.
Warning:
Clean the water filter located at the back of the unit at least once per month. Refer
to Maintenance Instructions.
Warning:
Use sterilization indicator test strips to check that sterilization has been
successful.
Warning:
Always check the water level in the reservoir before running a sterilization cycle. If
the LOW WATER indicator light illuminates, add distilled water. If the water is
sufficient, but the LOW WATER indicator light is still illuminating, refer to
Troubleshooting.
Warning:
Do not overfill the water reservoir. The water level must be maintained between
the green Full and Minimum labels on the right hand side of the sterilizer.
Warning:
The ADD WATER indicator light (Figure 16 & 18) will illuminate during the
sterilization cycle. This is part of normal operation and no action on the part of the
Содержание SA-230FA
Страница 6: ...5 damage to or destruction of the equipment or other property ...
Страница 27: ...26 Figure 26 Warning Please follow above Figure and place wrap vertically inside the sterilization box ...
Страница 36: ...35 Model SA 230FA SA 260FA Additional controls Start Reset Vacuum Pressure release ...
Страница 38: ...422 02014 02 ...