Electrical fast
transient/
burst IEC
6100-4-4
±2 kV for
power
supply lines,
±1 kV for
input/output
lines
Not
applicable
Main power quality
should be that of a
typical hospital or
commercial environment.
Surge IEC
6100-4-5
±1 kV
differential
mode, ±2
kV common
mode
Not
applicable
Main power quality
should be that of a
typical hospital or
commercial environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 6100-4-11
<5% UT
(>95% dip in
UT) for 0.5
cycle, 40%
UT (60% dip
in UT) for 5
cycles, 70%
UT (30% dip
in UT) for
25 cycles,
<5%UT (95%
dip in UT) for
5 seconds
Not
applicable
Main power quality
should be that of a
typical hospital or
commercial environment.
Power
frequency
(50/60Hz)
magnetic
field IEC
6100-4-8
3 A/m
3A/m
Power frequency
magnetic fields should
be that of a typical
hospital or commercial
environment.
NOTE: UT is the AC main voltage prior to application of the test level.