Technical Specifications
Specification
Technical Description
Channels
Dual; isolated between channels
Power Supply
One 9V battery or optional AC power
adaptor
Operating
Conditions
Temperature: 40°F - 104°F
Relative Humidity: 30% - 75%
Atmospheric Pressure: 525 mmHg -
795 mmHg
Storage Conditions
Temperature: 14°F - 122°F
Relative Humidity: 10% - 90%
Atmospheric Pressure: 525 mmHg -
795 mmHg
Tolerance
There may be a ±10% tolerance of
output intensity and ±5% on all other
setting parameters
Electrode Detection
Function
The amplitude level will be reset to
0mA when the amplitude level is
12mA or greater and an open circuit
is detected at either channel
Waveform
Mono-phase square pulse wave
Timer
15 minutes, 30 minutes, 60 minutes
and continuous
Pulse Amplitude
Adjustable, 0~105mA at 1,000 ohm
load at each channel; 1mA/step
Pulse Width
From 100 to 260 µs
Pulse Rate
From 50 to 150 Hz
Size
4.5"L x 2.55"W x 0.9"H
Weight
4.48 oz. (battery included)