10
Specification Technical Description
Channels
Dual, isolated between channels.
Pulse
Amplitude
Adjustable, 0-100 mA peak into 500 ohm load each
channel.
Pulse Rate
Adjustable from 2 to 150 Hz.
Pulse Width
Adjustable from 50 to 300 µs.
Modes
B(Burst), N(Normal), M(Modulation), SD1(Strength
Duration), SD2
Normal
Mode
The pulse rate and pulse width are adjustable. It
generates continuous stimulation based on the
setting value.
Burst Mode
Burst rate: Adjustable, 0.5-5 Hz
Pulse width: Adjustable, 50-300 µs
Frequency: Fixed at 100 Hz
Modulation
Mode
Modulation mode is a combination of pulse rate
and pulse width modulation. The pulse rate and
width are automatically varied in a cycle pattern.
The pulse width is decreased by 50% from its
original setting in 0.5 seconds, and then the pulse
rate is decreased by 50% from its original setting
in 0.5 seconds. Total cycle time is 1 second. In this
mode, pulse rate (2-150Hz) and pulse width (50-300
µs) are fully adjustable.
SD1 Mode
The SD1 (Strength-Duration) mode consists of
automatic modulation intensity and pulse in a 40%
range. The intensity is always increasing while
the pulse width is decreasing and vice versa. The
intensity is decreased by 40% while the pulse
width is increased by 40% in 5 seconds. In the next
5 seconds, the intensity is increased by 40% while
the pulse width is increased by 40%. Total cycle
time is 10 seconds. Pulse rate (2-150Hz) and pulse
width (50-300 µs) are fully adjustable.
Technical Specifications