background image

 

IonPac TCC-LP1, TCC-ULP1, and TCC-XLP1 
Product Manual 

Page 16 of 17 

 

 

 

 

 

 

Doc. No. 034973-06 

©2005 DIONEX 

September 2005 

 

5.2.

 

High Background, Noise, or Baseline Instability 

Normally, problems such as high background, noise, or baseline instability will not be attributable to the TCC-LP1, TCC-
ULP1, or TCC-XLP1 column.  These problems usually originate in either the analytical column or the post-column detection 
chemistry.  Before checking the TCC-LP1, TCC-ULP1, or TCC-XLP1 as the source of system background noise, consult the 
appropriate troubleshooting sections in the analytical column's Product Manual, the Ion Chromatograph Operator’s Manual, 
and the detector manual. 
 
If the source of the high background noise is isolated to the TCC-LP1, TCC-ULP1, or TCC-XLP1, then proceed with the 
following steps. 
 

A.

 

Be sure that the TCC column is not leaking. 

 

B.

 

Be sure that the eluents are correctly formulated. 

 

C.

 

If you are using an Cation Micro Membrane Suppressor III (CMMS® III), be sure that the regenerant is formulated 
correctly. 

 

D.

 

Be sure that the eluents are made from chemicals with the recommended purity (see Section 3, “Operation”). 

 

E.

 

Be sure that the deionized water used to prepare the reagents has a specific resistance of 18.2 megohm-cm. 

 

F.

 

Be sure that the Cation Self-Regenerating Suppressor (CSRS® ULTRA II), the Cation Micro Membrane Suppressor 
(CMMS III), or the Cation Atlas® Electrolytic Suppressor (CAES) is suppressing correctly by bypassing the TCC 
column and making direct injections. 

 
 

5.3.

 

Poor Peak Shape 

In some instances, poor peak shape in Ion Chromatography may be caused by a contaminated TCC column.  To clean the 
TCC Column, see, “Column Cleanup of Polyvalent Cations and Acid-Soluble Contaminants” in the Column Care Appendix. 
 
When pursuing pre-concentration with a pump, ensure that the pump has a pulse damper installed.  Failure to dampen the 
pump pulsations may result in damage to the TCC columns. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Assistance is available for any problem during the shipment or operation of Dionex instrumentation, columns, and 
consumables through the Dionex North America Technical Call Center at 1-800-DIONEX-0 (1-800-346-6390) or 
through any of the Dionex Offices listed in “Dionex Worldwide Offices” on the Dionex Reference Library CD-ROM. 

 

Содержание IonPac TCC Series

Страница 1: ...IonPac TCC...

Страница 2: ...ptember 2005 PRODUCT MANUAL FOR TRACE CATION CONCENTRATOR TCC TCC LP1 Column Low Pressure 4 x 35 mm P N 063079 TCC ULP1 Column Ultra Low Pressure 5 x 23 mm P N 063783 TCC XLP1 Column Extremely Low Pre...

Страница 3: ...acity 10 4 EXAMPLE APPLICATIONS 13 4 1 Concentration in the Manual Mode 13 4 2 Concentration in the Manual Mode 14 5 TROUBLESHOOTING GUIDE 15 5 1 High Backpressure from a Contaminated Inlet Bed Suppor...

Страница 4: ...throughout this document to emphasize important points The symbols are shown below along with the purpose of the information SAFETY Safety information can help prevent bodily harm WARNING Warning inf...

Страница 5: ...CC ULP1 is less than 45 psi 0 31 MPa at 1 mL min and the TCC XLP1 is less than 30 psi 0 21 MPa at 1 mL min The large resin particle size 20 m in the TCC LP1 TCC ULP1 and TCC XLP1 makes it possible to...

Страница 6: ...ional Group Hydro phobicity TCC LP1 4 mm 20 55 260 eq col Carboxylic Acid Very low TCC ULP1 5 mm 20 55 260 eq col Carboxylic Acid Very low TCC XLP1 6 mm 20 55 260 eq col Carboxylic Acid Very low TABLE...

Страница 7: ...Column Description The IonPac Trace Cation Concentrator Low Pressure TCC LP1 4x35 mm and IonPac Trace Cation Concentrator Ultra Low Pressure TCC ULP1 5x23 mm columns consist of the following component...

Страница 8: ...ugh to ensure good quantification WARNING The flow direction during the concentration step is critical In order to ensure optimal system performance it is recommended that concentration always be perf...

Страница 9: ...IONEX September 2005 Figure 3 shows the configuration for sample loading using a Rheodyne valve FIGURE 3 Configuration for Determining Trace Levels of Cations Figure 4 shows the configuration for samp...

Страница 10: ...o 8 days Recommended storage vessels are Corning tissue culture flasks The following procedure should be used for storage of g L level samples A Rinse the polystyrene container and cap twice with deio...

Страница 11: ...he analytes on the ion exchange sites and result in their elution to waste during the loading process B Conversely qualitative analysis of ions with higher affinities for the resin in the presence of...

Страница 12: ...e rate at which the sample is loaded C The pH of the sample D The ionic strength of the sample E The amount and capacity of resin in the column The breakthrough volume is determined as follows A Prepa...

Страница 13: ...2005 DIONEX September 2005 FIGURE 6 Linearity Determinations for Concentrator Injection Time min that the simulated sample is pumped through the TCC LP1 TCC ULP1 or TCC XLP1 after the single ion injec...

Страница 14: ...thanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL Temperature 35 C SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode MMS Regenerant 100 mN...

Страница 15: ...n Eluent 10 mM Methanesulfonic acid Eluent Flow Rate 1 0 mL min Injection Volume 1 0 mL SRS Suppressor Cation Self Regenerating Suppressor CSRS ULTRA AutoSuppression Recycle Mode or AES Suppressor Cat...

Страница 16: ...back on to the column Follow the instructions below to change the bed support assembly using one of the two spare bed support assemblies included in the ship kit provided with the column A Disconnect...

Страница 17: ...D Be sure that the eluents are made from chemicals with the recommended purity see Section 3 Operation E Be sure that the deionized water used to prepare the reagents has a specific resistance of 18...

Страница 18: ...he gradient pump Direct the effluent from the TCC LP1 TCC ULP1 or TCC XLP1 directly to a waste container C Set the flow rate to 1 mL min D Pump the 1 M HCl 0 1 M KCl solution through the column for 15...

Отзывы: