ASE 150 Operator’s Manual
34
Doc. 065207-02 9/08
3.1.4
Installing the Cell Filter
A disposable filter must be installed in the bottom end cap of the sample
cell before sample is loaded. The filter prevents blockage of the frit in the
bottom cap. For installation instructions, see
and
. Refer
to the information here to determine the appropriate filter material and
size.
Cell filters are available in two materials:
•
Use
cellulose
filters for most extraction methods that use organic
solvents.
•
Use
glass-fiber
filters for aqueous extractions, where cellulose may
provide inadequate filtration or may interfere with the analytical
technique.
Cell filters are available in two sizes:
•
Use 27-mm filters for 1
-
mL, 5
-
mL, 10
-
mL, or 22
-
mL cells.
•
Use 30-mm filters for 34
-
mL, 66
-
mL, or 100
-
mL cells.
Each ASE 150 Startup Kit includes a package of 100 cellulose filters, in
the size appropriate for the sample cell size to be used. For information
about reordering filters or ordering filters in a different material or size,
refer to
.
To install a 27-mm filter in a 1-, 5-, 10-, or 22-mL cell:
1. Unscrew the bottom cap from
the cell body.
To identify the cell top and
bottom, designate the end of the
cell with the groove around the
body and the Dionex logo, serial
number, and cell size as the cell
top (see
Figure 3-2. Sample Cell Top (Smaller Cells)
Groove
Logo, Serial
Number,
and Size
Содержание ASE 150
Страница 1: ...ASE 150 Accelerated Solvent Extractor Operator s Manual Document No 065207 Revision 02 September 2008...
Страница 32: ...ASE 150 Operator s Manual 24 Doc 065207 02 9 08...
Страница 64: ...ASE 150 Operator s Manual 56 Doc 065207 02 9 08...
Страница 76: ...ASE 150 Operator s Manual 68 Doc 065207 02 9 08...
Страница 114: ...ASE 150 Operator s Manual 106 Doc 065207 02 9 08...
Страница 132: ...ASE 150 Operator s Manual 124 Doc 065207 02 9 08...