12. Toggle the needle switch to the
UP
position and remove the collection bottle
containing the extract from the bottle holder. Refer to Section 3.5 for post-
extraction procedures.
3.5
Post-Extraction Procedures
3.5.1
Cleaning the Cells
After an extraction, empty the cells and rinse the cell bodies and end caps
with water or organic solvent. For most applications, it is sufficient to
simply rinse the end caps; however, if necessary, disassemble the end caps
and sonicate or soak in solvent to clean them. See Section 5.1 for
instructions on how to disassemble the cell.
The cell bodies (but not the caps) can be cleaned in a dishwasher or high-
temperature cleaning unit, provided that the temperature does not exceed
400 °C (752 °F).
3.5.2
Processing Extracts
The composition of the extracts generated by the ASE 100 is very close to
that generated by Soxhlet and other standard solid-liquid extraction
techniques when using the same solvent. Use the same analytical method
for ASE 100 extracts that you employed for extracts obtained from other
techniques.
Cells are extremely hot after an extraction. Be especially
cautious with cells that have been heated over 50 °C (122 °F).
Les cellules sont extrêmement chaudes après une extraction.
Laissez-les refroidir pendant au moins 15 minutes avant de les
manipuler. Faites particulièrement attention aux cellules qui
ont été chauffées à plus de 50 °C.
Die Zellen sind nach ihrer Entnahme sehr heiß. Lassen Sie die
Zellen mindestens 15 Minuten abkühlen. Seien Sie besonders
vorsichtig, wenn Zellen über 50°C erhitzt wurden.
Содержание ASE 100
Страница 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Страница 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Страница 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Страница 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Страница 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Страница 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Страница 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Страница 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Страница 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Страница 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...