background image

IonPacAS9-SC

034656-06

Page 23 of 28

and varying baseline will result. In this case the AMMS III or the ASRS-ULTRA will need to be serviced as outlined
in the AMMS III manual (see Document No. 031727, Product Manual for the Anion MicroMembrane Suppressor III,
the AMMS III or Document No. 031367, the “Product Manual for the Anion Self-Regenerating Suppressor-ULTRA,
the ASRS-ULTRA.”)

Carbonate Peak

When using a carbonate eluent, carbonate in the sample is not normally evident in the chromatogram. However, when
using other eluents (e.g., borate) a carbonate peak will be observed. The elution time of this peak will depend upon
the pH of the eluent. The higher the pH, the higher the carbonate/bicarbonate ratio. The divalent carbonate form tends
to elute later then the bicarbonate form. The converse is true as the pH is lowered. The borate eluent described in this
manual is an equimolar solution of sodium tetraborate and boric acid. The boric acid is added to lower the tetraborate
eluent pH. This places the carbonate peak between chloride and nitrite (see Figure 5, Anion Separation using 22 mM
Borate Eluent) when using a 22 mM borate eluent formulation. When using a more dilute eluent the carbonate is
retained longer since it is divalent. This explains why the carbonate peak elutes after the nitrite peak when using the
10 mM borate eluent formulation. Since the carbonate peak is reduced by almost two orders of magnitude during
passage through the AMMS III or the ASRS-ULTRA, this method is not useful for carbonate quantification.

Eluent Impurities

Use an Anion Trap Column (ATC-3, P/N 059660) to retard divalent anion contaminants (e.g., sulfate) in dilute eluents
(e.g., 10 mM borate) so that they do not concentrate on the analytical column when the dilute eluent is being run through
the column and then elute as sharp bands when the following strong eluent (e.g., 50 mM borate) is pumped through
the column. Dilute eluent contaminants eluting as sharp bands in the strong eluent can interfere with the quantification
of sample divalent anions (see Figure 6, Anion Separation using 10 mM Borate Eluent with Column Purge after
Nitrate). If the tetraborate salts or deionized water are contaminated (e.g., with sulfate) this will prolong the
reequilibration process from the strong eluent back to the dilute eluent.

6.2.3 A Contaminated Guard or Analytical Column

Remove the IonPac AG9-SC Guard and AS9-SC Analytical Columns from the system. If the background conductivity decreases,
the column(s) is (are) the cause of the high background conductivity. Clean or replace the AG9-SC at the first sign of column
performance degradation (compared to the original test chromatogram) to eliminate downtime. Clean the column(s) as instructed
in “Column Care.”

6.2.4 A Contaminated Anion Trap Column, ATC-3

When doing gradient analysis, has the Anion Trap Column,  the ATC-3 (4-mm) been installed correctly? If it has not, install one
as directed in Section 3.5, Installing the Anion Trap Column, and watch the background conductivity. If the background
conductivity is now low, this means that the ATC-3 is trapping contaminants from the eluent. The eluents probably have too many
impurities (see items 1 - 3 above).

If the ATC-3 is already installed, remove it. Is the background conductivity still high? If the background conductivity decreases,
the ATC-3 is the source of the high background conductivity.

A. Disconnect the ATC-3 (4-mm) from the injection valve and direct the outlet to waste.

B. Flush the ATC-3 with 200 mL of 70 mM Na

2

B

4

O

7

 at 2.0 mL/min on a 4-mm system.

C. Equilibrate the ATC-3 with the strongest eluent used during the gradient run. Use a flow rate of 2.0 mL/min on a 4-

mm system.

D. If the problem persists, replace the ATC-3.

Содержание 043185

Страница 1: ...for IonPac AG9 SC IonPac AS9 SC...

Страница 2: ...4656 06 Page 1 of 28 PRODUCT MANUAL for the IONPAC AG9 SC GUARD COLUMN 4 x 50 mm P N 043186 IONPAC AS9 SC ANALYTICAL COLUMN 4 x 250 mm P N 043185 DIONEX Corporation 2008 Document No 034656 Revision 06...

Страница 3: ...8 The Anion Atlas Electrolytic Suppressor AAES 8 3 9 The Anion MicroMembrane Suppressor AMMS III 8 3 10 Using AutoRegen with the ASRS ULTRA orthe AMMS III in the Chemical Suppression Mode 8 3 11 Using...

Страница 4: ...21 6 1 1 Finding the Source of High System Pressure 21 6 1 2 Replacing Column Bed Support Assemblies 21 6 2 High Background Or Noise 22 6 2 1 Preparation of Eluents 22 6 2 2 Borate Eluent Precautions...

Страница 5: ...e using standard operating conditions of at least 14 000 plates meter The IonPac AS9 SC can be operated at flow rates up to 3 0 mL min with eluents that have a pH between 2 and 11 Eluents may contain...

Страница 6: ...3 or recommended gradient mixers Pumps Use the GP40 GP50 IP20 IP25 in Standard Bore Configuration The GM 3 Gradient Mixer should be used for gradient analysis on systems other than the GP40 GP50 IP20...

Страница 7: ...DIONEX Worldwide Offices 3 2 The Sample Concentrator The Low Pressure Trace Anion Concentrator Column TAC LP1 P N 046026 the Trace Anion Concentrator Column TAC 2 P N 043101 the Anion MicroConcentrat...

Страница 8: ...ange resin which helps to minimize the baseline shift caused by increasing anionic contaminant levels in the eluent as the ionic concentration of the eluent is increased over the course of the gradien...

Страница 9: ...lytic Suppressor 3 9 The Anion MicroMembrane Suppressor AMMS III An Anion MicroMembrane Suppressor the AMMS III P N 056750 can also be used for applications that require suppressed conductivity detect...

Страница 10: ...e the Anion AutoRegen Regenerant Cartridge until it is completely expended and a sudden jump to very high background conductivity is observed 3 11 Using Displacement Chemical Regernation DCR with the...

Страница 11: ...uppressor 4 2 IonPac AS9 SC Operation Precautions CAUTION Filter and Degas Eluents Filter Samples Eluent pH between 2 and 11 and contains no hydroxide Sample pH between 2 and 13 3 mL min Maximum Flow...

Страница 12: ...ckground conductivity when using sodium carbonate bicarbonate isocratic or borate isocratic and gradient eluents with the AS9 SC columns The following table details the use of the above eluent types T...

Страница 13: ...eparation for Gradient Program 1 Eluent 10 mM H3 BO3 10 mM Na2 B4 O7 Thoroughly dissolve 3 814 g Na2 B4 O7 10 H2 O MW 381 42 g mole plus 0 618 g H3 BO3 MW 61 84 g mole in 700 mL degassed deionized wat...

Страница 14: ...vent concentration to be used during the analysis so that salts do not precipitate outin either the pump or the column Table 6 HPLC Solvents for Use with IonPac AS9 SC Columns Solvent Maximum Operatin...

Страница 15: ...2 5 orders of magnitude The concentrator column is used in lieu of the sample loop Pump the sample onto the concentrator column in the OPPOSITE direction of the eluent flow When using concentration t...

Страница 16: ...ChemicalPurityRequirements After running synthetic standards to calibrate your system you may find that real sample matrices foul your columns For this reason it is always advisable to use a guard col...

Страница 17: ...est Sample Loop Volume 25 L Analytical Column IonPac AS9 SC Analytical Column Eluent 2 0 mM Na2 CO3 0 75 mM NaHCO3 Eluent Flow Rate 2 0 mL min SRS Suppressor Anion Self Regenerating Suppressor ASRS UL...

Страница 18: ...tional resolution can be obtained at lower flow rates see Section 4 4 Resolution of Low Concentration Analytes Sample Loop Volume 25 L Analytical Column IonPac AS9 SC Analytical Column Eluent 1 8 mM N...

Страница 19: ...a2 CO3 1 7 mM NaHCO3 Eluent Flow Rate 1 0 mL min SRS Suppressor Anion Self Regenerating Suppressor ASRS ULTRA 4 mm AutoSuppression Recycle Mode or AES Suppressor Anion Atlas Electrolytic Suppressor AA...

Страница 20: ...22 mM Na2 B4 O7 Eluent Flow Rate 2 0 mL min SRS Suppressor Anion Self Regenerating Suppressor ASRS ULTRA AutoSuppression Recycle Mode or MMS Suppressor Anion MicroMembrane Suppressor AMMS III MMS Reg...

Страница 21: ...ytical Column IonPac AS9 SC Analytical Column Eluent 10 mM H3 BO3 10 mM Na2 B4 O7 with step change at 15 1 minutes to Purge 50 mM H3 BO3 50 mM Na2 B4 O7 Eluent Flow Rate 2 0 mL min SRS Suppressor Anio...

Страница 22: ...with particulates clogging the bed support a clogged High Pressure In Line Filter the suppressor or the detector cell To determine which part of the chromatographic system is causing the problem disco...

Страница 23: ...system is shown below ELUENT EXPECTED BACKGROUND CONDUCTIVITY 1 8 mM Na2 CO3 1 7 mM NaHCO3 14 18 S 5 mM Na2 B4 O7 3 4 S 10 mM Na2 B4 O7 4 6 S 22 mM Na2 B4 O7 5 7 S 6 2 1 Preparation of Eluents A Make...

Страница 24: ...ntisbeingrunthrough the column and then elute as sharp bands when the following strong eluent e g 50 mM borate is pumped through the column Dilute eluent contaminants eluting as sharp bands in the str...

Страница 25: ...sufficient to obtain the desired peak resolution or if the resulting increase in retention times is unacceptable clean the column see Column Care After cleaning the column reinstall it in the system...

Страница 26: ...rate In general the eluent flow rate for 4 mm applications it should be 1 0 mL min Refer to the Anion MicroMembrane Suppressor Product Manual Document No 034449 02 for assistance in determining that t...

Страница 27: ...ct Manual for column cleanup procedures C If you are running in the AutoSuppression External Water Mode turn off the external water and disconnect the external water line from the AAES REGEN IN port I...

Страница 28: ...eck to see if headspace has developed in the guard or analytical column e g due to improper use of the column such as submitting it to high pressures Remove the column s top end fitting see Section 5...

Страница 29: ...ice procedures Small baseline disturbances at the beginning or at the end of the chromatogram can be overlooked as long as they do not interfere with the quantification of the peaks of interest For DX...

Страница 30: ...mum of 10 minutes Cap both ends securely using the plugs supplied with the column COLUMN CLEANUP The following column cleanup protocols have been divided into three general isocratic protocols to remo...

Страница 31: ...guard column can be eluted onto the analytical column and irreversibly damage it If in doubt clean each column separately C Set the pump flow rate to 1 0 mL min D If your eluent contains a solvent th...

Страница 32: ...he outlet line of the AS9 SC Guard Column to a separate waste container CAUTION When cleaning an analytical column and a guard column in series ensure that the guard column is placed after the analyti...

Страница 33: ...uard Column to a separate waste container CAUTION When cleaning an analytical column and a guard column in series ensure that the guard column is placed after the analytical column in the eluent flow...

Отзывы: