Handheld Optical Power Meter
28
Input the corresponding optical power and click “calibration”, users can recalibrate the optical power by
themselves.
(2) Time setting
Fig. 6-28
Input the corresponding date and time, and then click “Set” to modify the time and date accordingly.
(3)Back to factory mode
6-29
Click “Reset”, and the unit will be returned to default setting (factory mode setting).
6.5.4 Close software
Click exit
,
pop up a window shown as below (figure 6-30):
6-30
6.6 Wavelength Automatic Identification
Connect power meter with a DINTEK Optical Light Source Unit.
Enable light source to be set under “Wave ID” operation mode: press
of light source to
emerge light output from light source. Hold down
for few seconds
,
light source will enter into
Wave ID mode, also “--AU” will be shown on the upper right of LCD for an indication.
To enable power meter under “Wave ID” operation mode: Hold down
for few seconds. The
power meter will be entered into Wave ID mode, also “--AU” will be shown on the upper right of LCD
for an indication.
Once the ID information is changed from light source (press
to change wavelength), after 3
to 5 seconds later, the detected information on optical power meter also will be changed
automatically according to light source. Please refer to below figure (6-31) for easy understanding.
Exit Wave ID mode: Hold down
again to exit Wave ID mode from power meter and hold
down
again to exit Wave ID mode from light source.