Cleaning and Maintenance
Copyright 2019 Dilon Technologies, Inc. All rights reserved. Version 1.7
Copyright 2019 Dilon Technologies, Inc. All rights reserved. Version 1.7
Inspect the stylet after cleaning. Discard if damage or
corrosion observed.
3.
STEP 2: Disinfect/Sterilize the CoPilot VL
®
+ Rigid Stylet
Clean and rinse the Rigid Stylet using an appropriate
surgical instrument detergent according to the
detergent manufacturer’s instructions.
2.
removed. Pay close attention to hard to reach areas.
Method Level Conditions Maximum
Cycles
Steris ®
Autoclave
(steam
cycle)
High
Steriliza-
tion
Standard cycles in the
following processors:
V-PRO 1 Standard Cycle;
V-PRO 1 Plus Lumen and
Non Lumen Cycles; V-PRO
maX Lumen, Non Lumen,
and Flexible Cycles; V-PRO
60 Lumen and Non Lumen
Cycles; and V-PRO maX 2
Lumen, Non Lumen, Flexible,
and Fast Non Lumen Cycles.
Minimum: 3 minutes
at 134° C (273° F) or 4
minutes and 132° (270°F)
Maximum: 8 minutes at
133°C (273°F)
100
100
Disinfection and Sterilization Methods: