30
Cleaning and Maintenance
Copyright 2019 Dilon Technologies, Inc. All rights reserved. Version 1.7
The CoPilot VL
®
+ Rigid Stylet should be High Level disinfected
between uses.
CoPilot VL
®
+ Rigid Stylet
cloth to remove the enzymatic cleaner.
STEP 4: Maintenance, Inspection, Testing, Storage
STEP 3: Drying
Inspect each component area for damage or
deterioration.
Allow components to air dry.
Package and store handle per facility policy to allow
device to remain clean, dry, and ready for service.
Check operational status.
1.
1.
3.
2.
Warning: Discard any component that
shows evidence of damage or deterioration.
Contact Dilon for component replacement.
STEP 1: Clean the CoPilot VL
®
+ Rigid Stylet
Rinse the stylet in clean tap water and scrub with a soft
bristled brush until all visible contamination has been
1.
With a new germicidal wipe, wipe to wet all exposed
surfaces of the handle and allow adequate contact
for disinfection as directed by the germicidal wipe
manufacturer.
4.
Содержание CoPilot VL+
Страница 2: ...2 Copyright 2019 Dilon Technologies Inc All rights reserved Version 1 7 ...
Страница 3: ...3 Copyright 2019 Dilon Technologies Inc All rights reserved Version 1 7 ...
Страница 4: ...Copyright 2019 Dilon Technologies Inc All rights reserved Version 1 7 4 ...
Страница 44: ...The first time Every time CoPilot VL ...