Digitimer Ltd - DS8R Operator’s Manual
Version 2.0
9
Digitimer Ltd.
Copyright ©2019
Explosion and Fire
Do not use the DS8R in an explosive atmosphere or in the presence of flammable anaesthetic gases.
High Voltages
Dangerous voltages are present within this unit. Do not remove covers, and refer servicing a
Digitimer representative.
Damage
Do not use the DS8R and/or any accessories if there are any signs of external damage.
Temperature
The DS8R and/or accessories should be stored or transported at temperatures between -40°C and
+70°C. The recommended usage temperature range is b10°C and 40°C.
Moisture
The DS8R and/or any accessories must not be used if any parts are wet or damp.
Cleaning Maintenance & Servicing
If the stimulator requires cleaning, it should be unplugged from the mains power supply and wiped
with a damp cloth.
This equipment does not require any regular maintenance but Digitimer recommends that all our
products designed for human use are returned to the factory for inspection and re-testing every 2
years. Please contact us for a reference number and instructions before despatching the unit.
As no parts of the DS8R are expected to age the only limitation to the service life is likely to be the
ongoing ability to replace electronic components that may randomly fail. The components chosen
are used conservatively and are all well-established parts so availability can be expected for many
years. However, guidance from the Biomedical Engineering Advisory Group of the South Australia
Department of Health gives a life expectancy of 10 years for neuromuscular stimulators, such as the
DS8R. This figure may be useful for lifespan planning purposes.
Before each use - The case and all leads should be inspected for any damage. The equipment (or the
lead) should be sent for repair if any damage is found.
For all servicing requirements please contact Digitimer, who can advise on best course of action.
Unpacking
After unpacking the DS8R stimulator from the shipping carton, please inspect each piece for any sign
of shipping damage. Please contact the carrier and your distributor, or Digitimer Limited,
immediately if there is any damage. Do NOT dispose of the shipping carton, as the carrier will want
to examine it in order to process a damage claim. Digitimer Limited and their distributors insure all
shipments to cover shipping damage.
It is also advisable to keep the shipping carton in the event that the instrument needs to be returned
for service.