![Digitimer D360R-4 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/digitimer/d360r-4/d360r-4_operators-manual_2500968019.webp)
Digitimer Ltd – D360R-4 Operator’s Manual
Version 1.0
19
Digitimer Ltd.
Copyright ©2021
Section 2 - Hardware
Overview
Tour of the Hardware
Front Panel Components
1.
Power LED (Green)
The power LED is illuminated when the D360R-4 is connected to a live power supply and switched on
via the rear panel On/Off switch.
2.
Error LED (Amber)
Normally extinguished, the error LED provides the operator with information that can indicate a
fault or communication problem. The LED normally flashes several times following Power ON and is
then unlit.
If the amber LED continues to flash or starts to flash during use, then this is likely to indicate that the
headstage is missing or there is a headstage communication problem, possibly due to cable damage.
Please check the headstage cable for damage and confirm the headstage is correctly connected.
If the error persists, contact Digitimer or our representative for further guidance.
If the headstage is not connected OR becomes disconnected, the D360R-4 disconnects itself from
the USB port. This means that the D360R-4 Client Software will reset and wait for the headstage to
be re-attached.