Recharge the battery immediately so that there is no damage to the battery or errors
in the weight readings.
4.7
Overload
-
Verify overload condition. The cargo cells can be damaged.
4.8
Under weight.
- Verify the spinning of the weight cell.
- Verify that there aren´t any objects blocking the scale or pulling it on the contrary direction.
In the case that the error persists, get in contact with Technical Assistance.
4.9
Blinking Display
-
If the display is blinking after a command on the keypad (tara or printing), there is a
pending command waiting for the platform’s stability or the minimum weight set the
tara or to print.
-
If the display is blinking after starting up the scale and keeps on blinking for an
extended period of time the instrument is not able to stabilize a zero reading. Check
for vibrations in the platform.
-
If the problem persists, seek technical assistance.
4.10
Display with some segments dimly lit
4.11
The Scale doesn’t turn on
Symptom: The start-up message may light up completely but shuts down immediately or lights
up and shuts down before the start-up is finished.
Description of error: When the battery is low, this
error message is shown every two minutes.
Description of Error: The weight is over 5 unit
divisions of the Maximum Load.
Weight on the scale over a long period of time, can
cause small deformations in the cargo cell and the
loss of the initial zero reading.
Description of Error: The weight is equal to or less
than -20% of the Maximum Load.
Low weight cell indication, due to signal noise and
traction on the contrary direction on the weight cell.
Symptom: Digits or LED’s keep blinking or are
dimly lit. Seek technical assistance.