DWL-5x000XY DATA SHEET
6.5.4.1. Detailed Calibration Steps :
Byte [n]
Command
Descriptions
Byte [0]
0x06
From Computer
Byte [1]
0x01
To Sensor 1
Byte [2]
0x0B
Calibration Mode
Byte [3]
0xA0
Acknowledgement to sensor module
Byte [4]~[11]
0x00
Not Applicable
Byte [n]
Command
Descriptions
Byte [0]
0x61
To Computer / From Sensor 1
Byte [1]
0x1B
Single Axis Position / Calibration Mode
Byte [2]
0x00
Calibration Feedback
Byte [3]
0x00
Calibration Count
Byte [4]
0x13
Temperature Value
Byte [5]
0x01
0x01: Calibration Step 1 ready
Byte [6]
0x88
Temperature Value
Byte [7]~[9]
0xDD
Fixed Dummy
Byte [10]
0xAA
CRC16 High
Byte [11]
0x93
CRC16 Low
Figure 28. Tilt Sensor Module Position of Calibration Step 1
1.
Please refer to Appendix 1 for proper instrument setup for calibration. Send the follow
-
ing command from computer to respective sensor to make sure the sensor is ready for
calibration mode:
<< (To sensor):
0x06, 0x01, 0x0B, 0xA0, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00
2. Wait for response from the sensor. Once the calibration mode is activated, last 4 bits of
Byte 2 will show 0xB and Byte 6 will show 0x01.
>> (From sensor) :
0x61, 0x1B, 0x00, 0x00, 0x13, 0x01, 0x88, 0xDD, 0xDD, 0xDD, 0x22, 0x93
3. Locate the sensor module as following Figure 28. The sensor is ready for user to perform
calibration step 1.
4. Send the following command from computer to respective sensor to trigger on the cali-
bration step 1:
<< (To sensor):
0x06, 0x01, 0x0B, 0xB0, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00, 0x00