Why isn’t the chart keeping time or running slow?

•  The chart may be hung up or restricted, possibly 

caused by a rip on the outer edge of the chart 
or the CHART HUB, or the chart may be caught 
between the arm and pen(s) arm platform. 

•  Incorrect chart installed for the selected chart 

speed. 

Why did the chart stop turning?

•  Chart hung up or restricted, (ripped chart) 
•  Unit may be “locked up”, this can be confirmed 

by pressing any of the buttons on the keypad, if 
the unit is locked up there will be no response 
to button presses and the Chart Recorder may 
appear to be working, but the readings won’t 
change, also the chart will not rotate. Remove 
power and battery, and then re-power. 

Why don’t the display and chart match?

•  DIP SWITCHES set for a specific range, but using 

chart for another range, or vice versa. 

•  Pen(s) not inserted on pen arm(s) all the way 
•  To adjust pen(s) to match chart see “PEN HOME 

Adjust” section of this manual.

Why does the display read PROB?

•  Check the sensor pod and KTC probe(s). Make 

sure it is properly plugged into the unit.

•  Is the unit locked up? This can be confirmed by 

pressing any of the buttons on the keypad, if 
the unit is locked up there will be no response 
to button presses and the unit may appear to be 
working, but the readings won’t change, also the 
chart will not rotate. Remove power and battery, 
and then re-power.

Why does calibration seem to be off? 

•  What is the tolerance of the unit it’s being 

compared to? It’s ok if the unit is within the sum 
of the two tolerances. 

•  Has an outside calibration house attempted 

calibration? It may not have been adjusted 
properly. 

•  Walk through calibration adjustment, found in 

the manual

Why won’t the battery back-up work?

•  Are there good batteries in the chart recorder?
•  Keep in mind that battery backup varies 

significantly depending on temperature, pen 
movement and chart rotation speed. 

•  If there was a momentary power outage (Brown 

out), the unit may not have had enough time to 
recognize this and switch to battery mode. (The 
Chart Recorder may lock up or shut off in this 
situation) This situation may also occur if the unit 
is plugged into an outlet that shares the circuit 
with other machinery that has phase motors or 
compressors that cycle periodically. As these 
other machines cycle, they momentarily have a 
high current draw, therefore drawing from the 
unit.

Why won’t the chart recorder respond to changes?

•  Is the unit locked up? This can be confirmed by 

pressing any of the buttons on the keypad, if 
the unit is locked up there will be no response 
to button presses and the unit may appear to be 
working, but the readings won’t change, also 
the chart will not rotate. Remove power and 
batteries, and then re-power. 

Why won’t the unit power up?

•  Remove the batteries and power adapter for a 

minute or two, this will reset the unit. The unit 
should respond when the adapter is plugged in.

Dickson warrants that this line of instruments will be 
free from defects in material and workmanship under 
normal use and service for a period of twelve months 
after delivery.

This warranty does not cover routine calibration and 
battery replacement. For Specifications and Technical 
Support go to DicksonData.com

Contact Customer Service at 630.543.3747 for a Return 
Authorization Number (RA) before returning any 
instrument. Please have the model and serial numbers 
and PO ready before calling.

Note: Dickson shall not be liable for consequential 
or incidental damages resulting from failure or 
malfunction of its products.

FAC

TROUBLESHOOTING

WARRANTY

FACTORY SERVICE AND RETURNS

6

7

To setup the KT8 recorder for your specific 
application, you might need to change some of the 
DIP SWITCHES. The DIP SWITCHES are located on the 
back of the unit. Use a pen or small screw driver to 
flip the switches.

Remember to install the correct chart to match the 
corresponding DIP SWITCH setting.

DIP SWITCH SETUP

RECORDING TIME

7 Day (all models)

#1 DOWN 

— 

#2 DOWN

24 Hour (all models)

#1 DOWN 

— 

#2 UP

31 Day (KT8P2/3, KT856)

#1 UP 

— 

#2 UP

RECORD IN °F OR °C

°F

#3 DOWN

°C

#3 UP

TEMPERATURE RANGE

Select the temperature range that matches the chart paper

0° to 100° F/C

#4 DOWN 

 #5 DOWN 

 #6 DOWN

-50° TO 50° F/C

#4 DOWN 

 #5 DOWN 

 #6 UP

0° TO 250° F/C

#4 DOWN 

 #5 UP 

— 

#6 UP

0° TO 500° F/C

#4 UP 

— 

#5 DOWN 

— 

#6 DOWN

0° to 150° F/C

#4 DOWN 

— 

#5 UP 

— 

#6 DOWN

-20° to 20° F/C

#4 UP 

— 

#5 UP 

— 

#6 UP

0° to 1000° F/C*

#4 UP 

— 

#5 DOWN 

— 

#6 UP

0° to 2000 F/C*

#4 UP 

— 

#5 UP 

— 

#6 DOWN

* A203 High Temp Probe required. Range not available on Model KT8P0.

1

2

3

4

5

6

7

8

Slide toggle UP for ON position

Dip Switches 7 and 8 are not active on this recorder

Slide toggle DOWN for OFF position

Содержание KT8

Страница 1: ...rument models listed No other Instrument will work with the Replaceable Sensors listed The Replaceable Sensor must be connected to the Instrument for the Instrument to operate properly Calibration Only the Replaceable Sensor is calibrated The sensor and all calibration defaults and adjustments are stored on the Replaceable Sensor Accuracy is strictly controlled by the sensor no adjustments are mad...

Страница 2: ... solid green While in this state the ADJUST UP button will move the chart backward counter clockwise and the ADJUST DOWN button will move the chart forward clockwise Rotate the chart until the correct hour and day if applicable on the chart is referenced to the timing arrow Once you have set your chart press the PEN HOME button to exit Chart Adjust Mode The unit will take one minute to exit the Ch...

Страница 3: ...will not turn off 2 To raise the unit of measurement press the ADJUST DOWN button To lower the unit of measurement press the ADJUST UP button Pressing PEN HOME switches between the red and blue pens on a two pen unit and stores the current adjustment value 3 When Calibration is complete press the ON OFF button The adjustment is stored in memory even after you turn the unit off or ifAC power fails ...

Страница 4: ...has phase motors or compressors that cycle periodically As these other machines cycle they momentarily have a high current draw therefore drawing from the unit Whywon t the chart recorder respond to changes Is the unit locked up This can be confirmed by pressing any of the buttons on the keypad if the unit is locked up there will be no response to button presses and the unit may appear to be worki...

Отзывы: