We recommend that damaged instruments be
returned to our factory for complete repair. Before
returning any instrument or package, please call for
a Return Authorization (RA) number.
When calling, please have the following information
available:
1. P.O. number for out-of-warranty units.
2. Serial number of unit.
3. Model number of unit.
4. The problem you're experiencing with unit.
Clearly mark the RA number on the outside
of the package and return the instrument,
shipping prepaid. Delivery will be refused for
packages sent without RA numbers.
Dickson warrants the ET6 line of instruments will be
free from defects in material and workmanship or
accidental damage for a period of twelve months
after delivery. In the event of a claim under this
warranty, the product or part must be returned
to the factory for repair or replacement (shipping
prepaid) with a Return Authorization Number (See
repair information above). It will be repaired or
replaced at Dickson’s option without charge.
This warranty
does not
cover routine calibration,
pen, chart and battery replacement.
The foregoing warranty and remedy are exclusive
and in lieu of all other warranties either expressed
or implied. Dickson shall not be liable for conse-
quential or incidental damages resulting from failure
or malfunction of its products. Dickson makes no
warranty for products not manufactured by it or
for any products modified by buyer, or subject to
misuse or neglect.
Symptom Cause Check/Remedy
Pen trace too Tip too sharp, cartridge Sand tip, moisten or replace pen.
fine or absent. dried out.
Readings of pen and Wrong Chart CHECK DIP SWITCH #4,5,6
display do not Pen not on properly Push pen completely onto pen arm.
match exactly. Line thickness Check to see that proper chart is
installed.
Und
Below range of recorder Check temperature range
Out of calibration Harsh environments, See Calibration Procedure on p.8
or
stressful conditions,
or
Questionable accuracy or time. return to factory for recalibration.
Ovr
Above range of recorder Check temperature range
Prb Sensor missing Connect transducer to back of the
recorder (See diagram on pg. 3)
Charts (60/pack)
ET6 Ranges 24-hour Chart 7-day Chart
0 to +100 C658 C657
0 to +200 C672 C671
0 to +300 C674 C673
-10 to +10 C676 C675
-50 to +50 C652 C651
0 to 14 C668 C669
-20 to +120 C668 C669
NOTE:
(ET6-2pen only) Chart number C669 has the following 2 ranges imprinted on it
so you can record both pH and temperature simultaneously: 0-14 and (-20) to 120
Accessories/Probes
Pens (6 red) P222
Pens (3 red & 3 blue) P246
Software and cable A015
3-Pt. NIST traceable certificate of calibration N300
One year extended warranty E100
Replacement Parts
8
Troubleshooting
Factory Service and Returns
The Dickson Warranty
Dickson manufactures a complete line of temperature, humidity, air quality &
pressure instruments. Send or call for a free catalog of Dickson products.