OPERATOR’S MANUAL
GAC® 2100 Grain Analysis Computer (Blue, Gray, Agri Models)
11001-1688-201609
60 INSTRUMENT SETUP
stroke. Advance until the correct grain displays. If the end of the list is
reached, the list wraps to the beginning and starts again.
5. When the correct grain is found, press the
Enter
key. The final screen
appears for editing the TEST WEIGHT BIAS OFFSET value. The
adjustment consists of three settings– the whole number, tenths and
hundredths.
6. Press the
Left/Right Arrow
keys to select the desired digit to change.
7. Change the selected value with the
Up/Down Arrow
keys. The value
of the number is monitored on the TEST WEIGHT BIAS
ADJUSTMENT line toward the bottom of the screen.
8. When finished, press the
Enter
key to accept the final bias value.
– An asterisk appears next to the value to indicate the number is
recorded and a test weight bias has been applied.
9. Press the
Previous Page
key to return to the
Setup
menu.
NOTE:
When a grain measurement is
taken, an asterisk (*) appears in
front of the test weight value to
indicate a test weight bias is
enabled.
Figure 49
Setting The Test Weight Adjustment
1
3
&
5
0
&
SETUP MENU
1: CLEAR COUNTER
2: ID OPTIONS
3: OUTPUT OPTIONS
4: EDIT ACCESS CODE
SELECT ITEM NUMBER
<more>
5: DATE/TIME MENU
6: LABELS MENU
7: COM MENU
8: SERVICE MENU
SETUP MENU
1: QUICK KEYS MENU
2: TEST WEIGHT BIAS ADJUSTMENT
3: TEST WEIGHT SLOPE ADJUSTMENT
4: MOISTURE BIAS ADJUSTMENT
5: MOISTURE SLOPE ADJUSTMENT
SELECT ITEM NUMBER
(NEXT PAGE: STATUS)
TEST WEIGHT ADJUSTMENT
TEST WEIGHT ADJUSTMENT
GRAIN SELECTED: CORN
TEST WEIGHT BIAS: 0.00 LB/BU
1
TEST WEIGHT ADJUSTMENT
(9.9 <-->-9.9)
RESOLUTION (LB/BU)
0.01 0.10
1.00
CORN
TEST WEIGHT BIAS: 0.02 *
/
to SELECT to ENTER
/
to SELECT to SET
/ to ADJUST
ENTER ACCESS CODE:
Blue model
Left/Right Arrow Keys
Up/Down Arrow Keys
Previous Page Key