OPERATOR’S MANUAL
GAC® 2100 Grain Analysis Computer (Blue, Gray, Agri Models)
11001-1688-201609
CALIBRATE PROCEDURES 37
After Printing Audit Memory:
1. Press the
Previous Page
key.
2. After printing, select (1 key) to clear audit memory or (2 key) to retain
audit memory.
Figure 29
Printing Complete Screen (Blue Model)
To Clear Audit Memory:
Clearing Audit Memory should only be performed by an authorized user.
1. Press the (1 key) Clear Audit Memory on the display.
2. Locate the access port door located on the back of the instrument and
remove seal, if necessary, from the door.
3. Remove screw to open access port door.
4. Press the white security button located on the bottom of the circuit
board to advance to the next screen.
5. When the white push button is pressed, audit memory is cleared and
the unit returns to the Calibration menu.
PRINTING AUDIT MEMORY COMPLETE
1: CLEAR AUDIT MEMORY
2: DO NOT CLEAR AUDIT MEMORY
TO PROCEED
Unseal Access Port on Rear Panel
and Depress White Push Button
Previous Page Key