If a new patient is selected, the results screen for the other patient in the same card will appear;
if we select end test, other wells to be assessed are delayed for another study; and when the
same card is located on a future date and when the same card is put on a future date, "test 2"
warning message will be given by the system and the device directly interprets the second test.
This is shown on the picture below.
Picture 7:
Across System Gel Card Reader Read Card window
•
Edit
After the recording is completed, if any parts for editing will appear, we can edit those
by using the edit button. The registered patient on the patient list will be marked from the check
box. When we click on edit button, a password screen will appear and we will type "diapro" on
the password area; after making the necessary editing, recording will be completed. Editing
screen is shown below.
Card Barkode
Patient Name Last Name/Barkode
Donor Name Last Name/Barkode
Back
Result
Save
Reference: 900160 Revision Number: 01 Issue Date 16.03.2017