Picture 8: 

Across System Gel Card Reader Read Card window

 

All of the areas shown on the picture can be changed. When there is a change in the part 
indicating the well values, this will be done during the record and in the editing part and "M" will 
be assigned in front of the value. This is to indicate that there was a manual intervention. 

 

 

 

Lis Send

 

After the patient recording is done, these may be sent to lis. There are two ways to make this 
transmission. In order to determine the Lis send status; one should enter service menu, system 
settings and select "Lis Status" and/or "Save+Lis" on the window opened. The active status of 
both will not cause any problems. Let us examine the send statuses below. 

 

1-

 

Save+Lis  

 This sending type provides automated transmission to Lis after the patient is recorded. 

2-

 

Lis Status

 

Card Barkode 

Patient Name Last Name/Barkode 

Donor Name Last Name/Barkode 

Back 

Result 

Save 

Print 

Reference: 900160 Revision Number: 01 Issue Date 16.03.2017

10

Содержание Across System

Страница 1: ...CARD READER USER MANUAL Version 2 03...

Страница 2: ...ACROSS SYSTEM GEL CARD READER Version 2 0 Dia Pro T bbi r nler San Trade A www diapro com tr Reference 900160 Revision Number 01 Issue Date 16 03 2017 1...

Страница 3: ...ed for reading and assessing the agglutination reactions occur in Across gel cards In order for Across System Gel Card Reader to operate properly original software and image processing card should be...

Страница 4: ...r transmitting the view on the screen of device to an external monitor USB inserts are used for connecting computer peripheral units such as Mouse Keyboard printer Ethernet insert is used to share the...

Страница 5: ...hen the power cable or connector is damaged Do not use the device in places which are not appropriate for its design or with improper material Do not use any other accessories than those designed or r...

Страница 6: ...e Reader Printer Flash memory Accompanied by Authorized Technician or under his knowledge for backup 2 Use of Across System Gel Card Reader Click on the gel reader symbol on desktop to start using The...

Страница 7: ...l see a window as in below Picture 4 Across System Gel Card Reader User Window The functions of buttons located in the user window picture above are as follows 1 After the card is placed into the devi...

Страница 8: ...rd installation will be done After the card is placed as in the picture with its bar code on the left the slide should be slightly pushed manually until the stopping point System automatically perceiv...

Страница 9: ...es when more than one patient is studied on the same card the screen below will appear after the recording is completed Picture 6 Across System Gel Card Reader test ending window Card Barkode Patient...

Страница 10: ...below Picture 7 Across System Gel Card Reader Read Card window Edit After the recording is completed if any parts for editing will appear we can edit those by using the edit button The registered pat...

Страница 11: ...these may be sent to lis There are two ways to make this transmission In order to determine the Lis send status one should enter service menu system settings and select Lis Status and or Save Lis on...

Страница 12: ...he screen can be printed out by clicking on the printing button located at the second row on the top left corner of screen Sample image can be seen on the picture below If you need a gel card imaged r...

Страница 13: ...s System Gel Card Reader Service Menu In order to access the Service Menu service part should be selected while entering the system In order to access the service menu diapro teknik password should be...

Страница 14: ...t made we will need to log in the system without any user name and password When we enter the user definition menu the window in the below will appear IDENTIFIYING USER CAMERA SETTINGS LOADING TEST CA...

Страница 15: ...ter the test loading menu two areas as active and passive tests will appear as indicated in the picture below Reference 900160 Revision Number 01 Issue Date 16 03 2017 Picture 13 Across System Gel Car...

Страница 16: ...contaminated by using 70 alcohol solution Cleaning and or decontamination transactions should be executed by using protective gloves The cleaning and or decontamination of device surface should be rea...