![diagenode Bioruptor Pico Скачать руководство пользователя страница 44](http://html.mh-extra.com/html/diagenode/bioruptor-pico/bioruptor-pico_user-manual_2492802044.webp)
44
BIORUPTOR
®
PICO
MANUAL
Technical data
General data
The device sound pressure level is below 70 dB. According to EC Directive 2006/42/EC the sound pressure level
of the devices is therefore not specified further.
SPECIFICATION
VALUE
UNIT
Placement
Indoor areas
Placement height above sea level
up to 2,000
m
Humidity
Maximum relative humidity 80%
Ambient temperature range
15 - 25
°C
IP protection rating
IP 32
Degree of soiling
2
Clearance from surroundings (front and
rear sides)
20
cm
Overvoltage
Overvoltage category II and transient overvoltages
according to category II
Protection class for electrical operating
equip- ment DIN EN 61 140 (VDE 0140-1)
1
Classification according to DIN 12 876-1
(class designation / identification)
I/NFL (non-flammable liquids, only)
Display
7-segment, LED
Display resolution
1
°C
Adjustment resolution
1
°C
Temperature stability
+- 2
K
Storage temperature range
5 - 40
°C
Transport temperature range
-20 - 60
°C
WORKING TEMPERATURE RANGE
DIMENSIONS (W X D X H)
WEIGHT
°C
mm x mm x mm
kg
BC 100
2 - 20
200 x 390 x 495
26,9