13
I F U
6. Technical Specifications
Control box
TiMotion, Voltage 220V-240VAC, 50Hz, 1250mA/24VDC, 70VA, IP66, Class II (Dou-
bled insulated)
Batteries
TiMotion, 24 V - 4A, 96Wh, Rechargeable Li-Ion Battery, IP66, Cable length 600mm,
Duty cycle: 10%, Max. 2min. ON/18 min./OFF
Actuator
12VDC, Max 8,0A, Push: 2000N, Self-locking capacity (min): 1000N, Stroke 400mm
+/-2 mm, IP66.
Duty cycle: 10%, Max. 2min.ON/18min. OFF
Hand Control
Cable length straight 300mm + coiled 400 mm, IP66
IP class
IPX5
Material
Steel
Castors
125/100/75mm
Weight
20 kg
Expected Service life
8 years
Max patient weight
240 kg / 529 Ibs
Operating environment
5 °C to 40°C, relative humidity 20% to 80% non-condensing, and atmospheric
pressure 700 to 1060 hPa.
Storage and transportation
environment
5 °C to 50 °C, relative humidity 20% to 80% non-condensing, and atmospheric.
Environment pressure 700 to 1060 hPa.
Height
125 mm wheels model minimum 900 mm and maximum 1300 mm
100 mm wheels model minimum 880 mm and maximum 1280 mm
75mm wheels model minimum 850 mm and maximum 1250 mm
Length
850 mm
Width
775 mm
Turning Width
125 mm wheels: 1158 mm, 100 mm wheels: 1130 mm, 75 mm wheels: 1130 mm
Materials in contact with
human skin
Materials that user and operator/caregiver are exposed to during usage of the pro-
duct for prolonged time:
Poroflex (synthetic leather cushions) P
x
Polyurethane (on PU cushions) P
Dryflex (handles) P
PA6 + 20% GF (brake lever) P
Hand control C
X
Applied parts:
Armrest P
Handles P
Knee pad P
Foot rest support P
Foot rest support clamps C
Handles adjusting knobs C
Hand controller and its holder. C
Handles to adjust the armrest width. C
Hand control C
Charger C
Charger cable C
Footplate
x
P - patient,
X
C - caregiver
Содержание SystemRoMedic 56-304
Страница 2: ...Table of contents English 4 Svenska 19 ...