3. OPERATION
Page 49
© 2004-2007 DH Instruments, a Fluke Company
When the run AutoZ selection is made, if a HEAD correction is currently active
(see Section 3.3.7) the head correction is momentarily disabled while running
AutoZ to avoid “zeroing out” the head value.
The value of P
offset
is always displayed and entered in Pascal (Pa).
Run AutoZ by Entry
AutoZ by entry allows the value of
P
std,0
(see Section 3.5.1,
PRINCIPLE) to be
entered directly from the RPM4 front panel. This provides a simple way of
AutoZeroing relative to an independent reference device such as a house
barometer that does not interface directly with RPM4.
To access run AutoZ by entry press
[AutoZ]
,
<1entry>
. The display is:
1. Real time pressure reading (without head
correction), pressure unit of measure and
measurement mode of the active Q-RPT.
2.
Active Q-RPT position designator.
3. Entry field for the value of
P
std,0
. in the current
pressure unit of measure.
96.7755 kPa a Hi
Pstd,0:0.00000
Enter the value of the AutoZ reference (
P
std,0
) in the same unit of measure as the
top line display and press
[ENT]
. RPM4 logs the reading and calculates a new
AutoZ offset value. The next display is:
Press
[ENT]
to activate the new value of
P
offset
or
[ESC]
to start over with entry of a
new AutoZ reference (
P
std,0
) value.
Old P
offset
: 0.0 Pa
New P
offset
: 8.3 Pa
The value of P
offset
is always in Pascal (Pa). The value of P
std,0
is entered in
the current pressure unit of measure.
Run AutoZ by COM2
AutoZ by COM2 allows the value of
P
std,0
(see Section 3.5.1
Principle) to be read
automatically from a
DHI
RPMx connected to the RPM4 COM2 communications
port. The RPMx is read and the new
P
offset
is calculated automatically.
Содержание RPM4 - VERSION 2004
Страница 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 114 N NO OT TE ES S...
Страница 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 138 N NO OT TE ES S...