RPM4™ OPERATION AND MAINTENANCE MANUAL
© 2004-2007 DH Instruments, a Fluke Company
Page 48
3.3.9.2
[AUTOZ] IN ABSOLUTE MODE
PURPOSE
To “rezero” the active Q-RPT in absolute measurement mode
See Section 3.5.1,
Principle for a full explanation of the AutoZ function.
OPERATION
For the AutoZ function key to run AutoZ, AutoZ must be turned ON for the
active Q-RPT and measurement mode. AutoZ ON is indicated by <z> to the
left of the range designator on the first line of the main run screen. AutoZ
ON and OFF is set using [SPECIAL], <1AutoZ> (see Section 3.5.1). If AutoZ
is OFF for the active range, and measurement mode, <AutoZ OFF> is
displayed when [AutoZ] is pressed.
Running AutoZ in or parallel mode (see Section 3.5.7.3) autozeros both the
Hi and Lo Q-RPTs simultaneously.
To run AutoZ in absolute measurement mode, set absolute measurement mode
(see Section 3.3.3) and press
[AutoZ]
from the main run screen. The display is:
1.
Active Q-RPT designator.
2.
Current measurement mode (absolute).
3.
Selection of source of
P
std,0
reference to which
to AutoZ.
AutoZ Hi abs by:
1Entry 2COM2 3Lo RPT
Selecting
<1Entry>
allows the value of
P
std,0
to be entered from the front panel keypad.
Selecting
<2COM2>
allows the value of
P
std,0
to be read automatically from another
DHI
RPMx connected to RPM4’s COM2 communications port.
Selecting
<3Lo RPT>
is shown only if the active Q-RPT is Hi and there is a Lo
Axxx Q-RPT in the RPM4. This selection allows the value of
P
std,0
for the Hi
Q-RPT to be read automatically from the RPM4s Lo Q-RPT.
Allow the RPM4 to stabilize at atmospheric pressure and ambient
temperature for 10 to 15 minutes before running AutoZ in absolute mode.
If running AutoZ results in a value of P
offset
that is greater than ± 0.005 % FS of
the span of the Q-RPT that is being AutoZeroed, the Q-RPT and/or the reference
used as the source of P
std,0
may be out of tolerance or the AutoZ process may
have been faulty. Before activating a new P
offset
greater than ± 0.005 % FS of
the active Q-RPT, check to be sure that both the Q-RPT and the reference
were in good working order, properly vented to stable atmospheric pressure,
at the same height, and reading in the same pressure units when AutoZ was run.
Содержание RPM4 - VERSION 2004
Страница 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 114 N NO OT TE ES S...
Страница 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 138 N NO OT TE ES S...