3. OPERATION
Page 45
© 2004-2007 DH Instruments, a Fluke Company
Use
[SDS]
to OPEN SDS if it is CLOSED for the active Q-RPT and/or to CLOSE SDS for the
active and inactive Q-RPT.
When
[SDS]
is pressed, there are three possible results depending whether the active Q-RPT is
SDS equipped and ON (see Section 3.5.4.2), and whether its SDS is OPEN or CLOSED.
1. If the active
Q-RPT IS NOT SDS
equipped or its SDS if “full time OFF”
(see Section 3.5.4.2):
<No SDS on RPT>
or
<SDS is OFF and OPEN>
is displayed for 5 seconds and
operation returns to the main run screen. If there is a second Q-RPT with SDS, its SDS
is CLOSED.
2. If the active
RPT IS SDS EQUIPPED
and SDS on the active RPT is OPEN
:
SDS is CLOSED and operation returns to the main run screen.
3. If the active
RPT IS SDS EQUIPPED and SDS on the active RPT is CLOSED
:
The OPEN SDS routine is presented. The display is :
1. Maximum acceptable pressure of active Q-RPT (always
absolute on an Axxx Q-RPT and gauge on a Gxxx Q-RPT).
2. Position designator of Q-RPT for which SDS will be
opened.
Pmax! 350 kPa a Hi
Open SDS? 1no 2yes
Check that the pressure applied to the
TEST(+)
port does NOT exceed the maximum
acceptable value for the Q-RPT. Select
<1no>
to return to the main run screen without
OPENING SDS, select
<2yes>
to OPEN SDS and return to the main run screen with
SDS OPEN (the active Q-RPT open to the
TEST(+)
port).
Opening SDS with a pressure greater than Pmax! applied to the test port may cause
overpressure damage to the Q-RPT.
In differential and parallel measurement modes, opening SDS acts simultaneously on
BOTH THE Hi AND Lo Q-RPTs (see Section 3.3.8.1). The Pmax! pressure is the Lo RPT.
Be sure to check BOTH Q-RPT TEST(+) ports are at a safe pressure before opening SDS.
The [SDS] key is also the [2] key which is pressed to select <2Yes> to OPEN SDS.
Thus, SDS can be OPENED by two rapid presses of [SDS]. Use this feature to
conveniently OPEN SDS but always check that the pressure connected the TEST(+) port
does NOT exceed Pmax! before doing so.
When changing ranges causes a change of active Q-RPT, SDS is automatically CLOSED for
the Q-RPT that is being changed from, leaving the inactive Q-RPT with SDS CLOSED.
The Lo and Hi Q-RPTs’ SDS, if present, can be controlled directly regardless of which
Q-RPT is currently active (see Section 3.5.4).
SDS can be permanently opened so operation is as if SDS were not present
(see Section 3.5.4.2).
Содержание RPM4 - VERSION 2004
Страница 10: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page VIII N NO OT TE ES S...
Страница 16: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page VI N NO OT TE ES S...
Страница 124: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 114 N NO OT TE ES S...
Страница 144: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 134 N NO OT TE ES S...
Страница 148: ...RPM4 OPERATION AND MAINTENANCE MANUAL 2004 2007 DH Instruments a Fluke Company Page 138 N NO OT TE ES S...