MPC1-1000 and MPC1-3000 User’s Manual
3.4
SETTING NEGATIVE GAUGE PRESSURE (BELOW ATMOSPHERE)
3.4.1
DECREASING PRESSURE
n
Close test outlet valve (4).
o
Rotate outlet selector valve (6) to vacuum position.
For small pressure changes, it may be possible to reach the desired pressure using the VV
only, proceed to Step
p
below.
p
Open pressure equalization valve (3) (pull valve outward).
OPENING THIS VALVE MAY RESULT IN A SUDDEN CHANGE IN THE TEST PRESSURE.
Ensure conditions are such that this either does not happen or that its effects
are inconsequential.
q
Place VV piston (2) at appropriate position, see Section 3.1.3.
r
With a vacuum source connected to the vacuum port, open test outlet valve until the
approximate desired pressure is reached.
s
Close equalization valve (push valve inward).
t
Adjust VV knob until desired pressure is obtained (or until the piston floats when
connected to a piston gauge).
u
To continue decreasing pressures, repeat Step
p
or
t
shown above.
Page
9
© 2001 - 2007
DH Instruments, a Fluke Company
Содержание MPC1-1000
Страница 1: ...MPC1 1000 and MPC1 3000 User s Manual 2001 2007 DH Instruments a Fluke Company ...
Страница 8: ...MPC1 1000 and MPC1 3000 User s Manual 2001 2007 DH Instruments a Fluke Company Page 4 N NO OT TE ES S ...
Страница 16: ...MPC1 1000 and MPC1 3000 User s Manual 2001 2007 DH Instruments a Fluke Company Page 12 N NO OT TE ES S ...
Страница 18: ...MPC1 1000 and MPC1 3000 User s Manual 2001 2007 DH Instruments a Fluke Company Page 14 N NO OT TE ES S ...
Страница 26: ...MPC1 1000 and MPC1 3000 User s Manual 2001 2007 DH Instruments a Fluke Company Page 22 N NO OT TE ES S ...
Страница 30: ...MPC1 1000 and MPC1 3000 User s Manual 2001 2007 DH Instruments a Fluke Company Page 26 N NO OT TE ES S ...