MPC1-1000 and MPC1-3000 User’s Manual
3.3.2
DECREASING PRESSURE
For small pressure changes, it may be possible to reach the desired pressure using the VV
only, proceed to Step
r
below.
n
Open pressure equalization valve (3) (pull valve outward).
OPENING THIS VALVE MAY RESULT IN A SUDDEN CHANGE IN THE TEST PRESSURE.
Ensure conditions are such that this either does not happen or that its effects
are inconsequential.
o
Place VV piston (2) at appropriate position, see Section 3.1.3.
p
Open test outlet valve (4) until the approximate desired pressure is reached.
q
Close equalization valve (push valve inward).
r
Adjust VV knob until desired pressure is obtained (or until the piston floats when
connected to a piston gauge).
s
To continue decreasing pressures, repeat Step
n
or
p
shown above.
© 2001 - 2007
DH Instruments, a Fluke Company
Page 8