VW MiniLogger, 2018/8/24
8
Set Clock
Click on the “Set Clock” but-
ton to set the MiniLogger’s
clock. To synchronize the
MiniLogger clock to your
computer’s clock, click the
“Match Computer” button. If
you want to set a different
time, click in the date and
time fields, type in values, and click OK.
Note 1:
The date display format is controlled by the short date
setting in Windows (Control Panel>Regional Settings>Date).
Note 2:
Both the MiniLogger and the Manager program use
four-digit years internally.
Set Sensor
Click the Set Sensor button to enter the sensor serial number,
the sweep frequency, and sensor calibration factors.
Sensor
ID
:
Unique identifier established by the user. Cannot be
the same as the serial number.
Sensor Serial Number:
Sensor serial number from the factory.
When you retrieve data and want to save it, the Sensor SN entry
will be suggested as a file name.
Coefficients:
Choose a Calibration Model for the sensors. For
instance, standard borehole piezometers would use VW_ABC
to specify a VW sensor using ABC Calibration factors. Choose
Single Channel. Enter ABC factors as noted on the sensor cali-
bration sheet.
Use ABC / TI factors:
These factors are used to convert Hz read-
ings to kPa or psi. TI factors integrate temperature readings to
automatically correct for temperature effects. See the sensor cal-
ibration sheet to see how the factors are applied.