17.
Care and Maintenance
17.1
Cleaning and Disinfecting the Probe Tip
Keep the probe tip clean and disinfected to prevent patient-to-patient infection. After each
patient, wipe the probe with a Q-tip soaked in 70% isopropyl alcohol, and then immerse the
probe tip (the clear cone) for 10 minutes in 70% isopropyl alcohol. The tip should be rinsed in
sterile distilled water before using.
WARNING:
The probe should NEVER be autoclaved or subjected to intense heat. As a
general rule, the above cleaning instructions are sufficient to disinfect the probe in ordinary
use. Do not scratch or chip the conical probe tip, which makes contact with the cornea.
The following disinfectants were found to be compatible with the probe tip material:
Disinfectant
Concentration Tested*
Cavicide Solution
(10-20%) Isopropyl Alcohol and (1-5%) Ethylene
Glycol Monobutyl Ether
Cavicide Wipe
(10-20%) Isopropyl Alcohol and (1-5%) Ethylene
Glycol Monobutyl Ether
Cidex
2.55% (w/w) Glutaraldehyde
Cidex OPA
6.2% by (w/w) Ortho-Phthalaldehyde
(1,2 – benzenedicarboxaldehyde)
Isopropyl Alcohol
70% (v/v) Isopropyl Alcohol
Household Bleach
0.6% (w/w) Sodium Hypochlorite
Hydrogen Peroxide
3% (w/w) H
2
O
2
Milton
2% (w/w) Sodium Hypochlorite
* The concentrations listed in this table are the specific concentrations that were tested by DGH to ensure compatibility with the
probe tip material. DGH does not endorse or recommend the concentrations listed in the table above.
WARNING:
DGH makes no claims about the biological effectiveness as a disinfectant of
any of the products listed above. Furthermore, DGH makes no claims regarding the
effectiveness of any of these products for killing any known, or unknown, bacteria, virus,
or other micro-organisms. DGH only claims that these products, when used properly, will
not harm the transducer tip.
!
!
Содержание DGH 55B
Страница 2: ...Page 2 of 87 DGH 55B INS OMENG Rev 3 This page intentionally left blank ...
Страница 27: ...Page 27 of 87 DGH 55B INS OMENG Rev 3 Fig 10 1 1 Single Patient Mode ...
Страница 29: ...Page 29 of 87 DGH 55B INS OMENG Rev 3 Fig 10 2 1 Multi Patient Mode ...
Страница 84: ...Page 84 of 87 DGH 55B INS OMENG Rev 3 Image from Panasonic Corporation PAN1322 SPP User Manual Rev 1 3 ...
Страница 85: ...Page 85 of 87 DGH 55B INS OMENG Rev 3 Image from Panasonic Corporation PAN1322 SPP User Manual Rev 1 3 ...
Страница 86: ...Page 86 of 87 DGH 55B INS OMENG Rev 3 Image from Panasonic Corporation PAN1322 SPP User Manual Rev 1 3 ...
Страница 87: ...Page 87 of 87 DGH 55B INS OMENG Rev 3 Image from Panasonic Corporation PAN1322 SPP User Manual Rev 1 3 ...