DGH 55B-INS-OMENG Rev: 3
18.
Troubleshooting Guide
PROBLEM /
ERROR MESSAGE
POTENTIAL CAUSE
SOLUTION
Device will not turn on
Dead batteries
Replace batteries
See section 16
Device screen is blank
In Sleep Mode
Press the PWR button
See section 9.3
Screen is on, but will not take
measurement.
In Standby Mode (indicated
by blinking black cursor in
upper left-hand corner)
Press the PWR button
See section 9.2
‘Plug In Probe’ Message on
screen
Probe is wet or has a residue
on it
Dry probe
See section 17.1 and 7.4
‘Check Probe’ message on
startup.
Probe is wet or has a residue
on it
Dry probe
See section 17.1 and 7.3
Will not initiate CalBox mode
While device is off, hold
down the DEL key and press
the PWR key
Device starts in CalBox
mode
See section 8
Black boxes across the top half of
the screen
1. Bad battery contact
2. Dead batteries
1. Clean battery contacts
2. Replace batteries
See section 16
Slow measurement cycle
1. Dented/scratched probe
tip
2. ‘Auto Rep Delay’
configuration
1. Replace probe. Contact
DGH Technology.
See section 19
2. Adjust configuration
See section 13.2.3
Cannot send files to PC/printer
1. Bluetooth not enabled
2. Devices not paired
3. COM settings incorrect
1. Enable Bluetooth module
See section 14.1
2. Pair devices
See section 14.2 and 14.4
3. Adjust COM settings
See section 15.5.3
‘Remote Device Not Found’ error
message when attempting to send
measurements to the software
Pairing not established
Pair devices then add device
to the software’s device list.
See section 14.4 and 15.4
Cannot find correct device on the
Pachmate 2
when trying to
export/print measurements
Pairing not established
Clear all pairing and re-pair
with preferred device.
See section 14.9 regarding clearing
See section 14.3 and 14.4
regarding pairing
Содержание DGH 55B
Страница 2: ...Page 2 of 87 DGH 55B INS OMENG Rev 3 This page intentionally left blank ...
Страница 27: ...Page 27 of 87 DGH 55B INS OMENG Rev 3 Fig 10 1 1 Single Patient Mode ...
Страница 29: ...Page 29 of 87 DGH 55B INS OMENG Rev 3 Fig 10 2 1 Multi Patient Mode ...
Страница 84: ...Page 84 of 87 DGH 55B INS OMENG Rev 3 Image from Panasonic Corporation PAN1322 SPP User Manual Rev 1 3 ...
Страница 85: ...Page 85 of 87 DGH 55B INS OMENG Rev 3 Image from Panasonic Corporation PAN1322 SPP User Manual Rev 1 3 ...
Страница 86: ...Page 86 of 87 DGH 55B INS OMENG Rev 3 Image from Panasonic Corporation PAN1322 SPP User Manual Rev 1 3 ...
Страница 87: ...Page 87 of 87 DGH 55B INS OMENG Rev 3 Image from Panasonic Corporation PAN1322 SPP User Manual Rev 1 3 ...