8.3
Items To Check
8.3.1
Daily Inspection Items
When turning on product power, confirm if power on sound goes off, and normal
initial screen is displayed.
If exposure time setting are changed by using manipulation button, confirm if time
is changed normally.
When digital sensor is connected with main body, confirm if it is recognized
normally.
When performing X-RAY exposure, confirm if beeping sound goes off, and exposure
button turns to yellow.
When charging battery pack, check LED light of cradle to confirm normal operation.
8.3.2
Monthly Inspection Items
Confirm if battery pack use period is proper. Purchase new battery pack if use time
of initial use is reduced in half.
CHECK
Battery pack is consumables, and use time is reduced gradually
compared with purchasing time. If use time is shortened to more
than half compared with first use frequency time, replace with new
battery pack.