32
Rated maximum
output power of
transmitter
W
Separation distance according to fre-
quency of transmitter
m
150 kHz to
80 MHz
d = 1.2
√
P
80 MHz to
800 MHz
d = 1.2
√
P
800 MHz to
2.5 GHz
d = 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed
above, the recommended separation distance
d
in metres
(m) can be estimated using the equation applicable to the
frequency of the transmitter, where
P
is the maximum out-
put power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency
range applies
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
This RC-DEX may be interfered with by other equipment even if that
other equipment complies with CISPR emission requirements.