EMC/EMI Compliance
The RC-DEX complies with the following EMC/EMI standards.
Standard
Test type
Note
47 CFR Part
15, subpart C
RF emissions
USA Federal
Communications Commission
(FCC) requirements to
intentional radiators.
EN 303 330-2
V1.3.1
RF emissions
incl. Spurious
emission
EMC and radio spectrum
matters for Short Range
Devices in the frequency
range 9 kHz – 25 MHz
IEC 60601-1-
2:2007
*adapted
protocol
EMC emission
Immunity, RF
and ESD
Medical electrical equipment.
General requirements for
basic safety and essential
performance.
Electromagnetic
compatibility.
EN 301 489-3
V1.4.1
Immunity, RF
and ESD
Standard for Low Power
Transmitters in the frequency
range
9 kHz – 40 GHz
*The device was tested in only one orientation that represents
the longest length (or worst case scenario). This is acceptable be-
cause of the relative small size of the device compared to the
wavelength of the RF used in the test.
Important notice for prospective hearing aid users
Good health practice requires that a person with a hearing loss
have a medical evaluation by a licensed physician (preferably a
physician who specializes in diseases of the ear) before purchasing
a hearing aid. Licensed physicians who specialize in diseases of
the ear are often referred to as otolaryngologists, otologists, or
otorhinolaryngologists. The purpose of medical evaluation is to
assure that all medically treatable conditions that may affect hear-
ing are identified and treated before the hearing aid is purchased.
Содержание RC-DEX
Страница 1: ...USER INSTRUCTIONS RC DEX Remote control for Widex tinnitus devices...
Страница 6: ...6 Rear view 1 Battery cover 2 Eye for key ring string attachment 1 2...
Страница 7: ...7 ACCESSORIES 1 Battery 2 Key ring string 1 2...
Страница 39: ...39...
Страница 40: ...Printed by HTO 2012 12 9 514 0237 041 01 9 514 0128 041S 03 Manufacturer CIB285 1212...