background image

DEWETRON

data acquisition

system

RS-232/485

conversion on
DEWETRON-

motherboard

DAQx or PAD

module

RS-232

interface

RS-485

businterface

8

RS-232/485 interface

HSI/DAQP modules can be configured via

 

RS-485 interface, PAD modules require 

this interface for all data transfers.
For all DEWETRON systems, an internal 

RS-232/485 converter is available 

(standard with DEWE-800, -2000, -2500, -3000, -4000, -5000 series systems). This converter allows 

communication with HSI/DAQP and PAD modules.
To communicate with the modules, the RS-232 interface has to be set to the following parameters:

baud rate: 

9600 bps 

data bits: 

parity: 

no parity 

stop bits: 

handshake: 

not required

HSI/DAQP module configuration

1. Push button selection

All ranges and filters can be selected directly by pressing the push buttons on the module. Approx. 15 sec. 

after changing range and / or filter, the range and filter information is stored in an EEPROM. This procedure 

increases the livetime of the EEPROM.
The current input range setting is shown all the time by LED. To change the range just press 

RANGE 

button 

a few times until the required range is displayed.

To see the current filter setting just press the 

FILTER 

button once. The corresponding LED is flashing for 

approx. 3 seconds. Within this time, the filter can be selected by pressing the 

FILTER 

button again. Approx. 

3 seconds after the last key activity, the information will be stored, the LED stops flashing and shows the 

input range again.

CAUTION: Power loss during this time leaves the module in the former settings.

2. RS-232/485 programming

All ranges and filters can also be selected via RS-232/485 interface. All new DEWE-800, -2000, -2500,

-3000, -4000, -5000 series systems are prepared as a standard to work with HSI/DAQP modules.

The easiest way to change the configuration is to use the DEWEConfig software, which comes as a standard 

with the DEWETRON data acquisition system.
Detailed information about HSI/DAQP modules programming for customer applications is available in the 

DEWE-Modules Programmers Reference Manual

.

CAUTION: All range and filter changes which are done via RS-232/485 interface are not stored in the 

EEPROM of the HSI/DAQP modules! You have to store this information in a separat initialisation file to 

keep settings information for next system start!

PAD module communication

All PAD modules are only working through the RS-232/485 interface. All new DEWE-800, -2000, -2500, 

-3000, -4000, -5000 series systems are prepared as a standard to work with PAD modules. The easiest

way to change the configuration is to use the DEWEConfig software, which comes as a standard with the

DEWETRON data acquisition system.
Detailed information about PAD modules programming for customer applications is available in the 

DEWE-Modules Programmers Reference Manual

.

General Module Information

Содержание DAQP-ACC-A

Страница 1: ...l supply you with accurate complete and reproducible measurement results for your decision making Look forward to the easy handling and the flexible and modular use of your DEWETRON product and draw u...

Страница 2: ...protected by copyright All rights are reserved Reproduction adaptation or translation without prior written permission is prohibited except as allowed under the copyright laws All trademarks and regis...

Страница 3: ...se of the information in this document Warranty Information A copy of the specific warranty terms applicable to your DEWETRON product and replacement parts can be obtained from your local sales and se...

Страница 4: ...wing general safety precautions must be observed during all phases of operation service and repair of this product Failure to comply with these precautions or with specific warnings elsewhere in this...

Страница 5: ...ained Keep away from live circuits Operating personnel must not remove equipment covers or shields Procedures involving the removal of covers or shields are for use by service trained personnel only U...

Страница 6: ...com The telephone hotline is available Monday to Friday between 08 00 and 12 00 CET GMT 1 00 and Monday to Thursday between 13 00 and 17 00 CET For the Americas please contact DEWETRON Inc 2850 South...

Страница 7: ...AQ series modules available The CAL KIT contains the required cables software and instructions General module specifications Module dimensions 20 x 65 x 105 mm 0 79 x 2 56 x 4 13 in W x H x D without...

Страница 8: ...e FILTER button again Approx 3 seconds after the last key activity the information will be stored the LED stops flashing and shows the input range again CAUTION Power loss during this time leaves the...

Страница 9: ...lected highpass filter to 300 kHz 2 to 5 dB fg Filters highpass 0 5 Hz filter 5 Hz filter 0 5 Hz and 5 Hz software selection 0 32 s time constant 0 032 s time constant Filters lowpass 1 kHz 10 kHz 100...

Страница 10: ...he GAIN button several times until the LED displays the desired input range Filter button Push the FILTER button once the LEDs will flash for approx 3 seconds and display the current filter setting Pu...

Страница 11: ...ber States relating to electrical equipment designed for use within certain voltage limits amended by the directive 93 68 EEC 89 336 EEC Directive on the approximation of the laws of the Member States...

Страница 12: ...C2 Notes...

Отзывы: