background image

DS-CAM-600

2Technical data

2.3 Resolutions / Framerates

HINT

Results   (FPS)   should   be   considered   only   as   approximate   MAX   values.   It  
depends on the system performance (CPU, JPEG compression, harddisk). We  
recommend to test cameras with your setup at about 0.8*max_FPS and higher  
to find where corrupted image, frame lost or buffer overrun will occur. 

Page 7/17

www.dewesoft.com

Document-Version:  1.0.0

Sensor type

CMV2000

Active pixels

2048 x 1088

Resolution

Active pixels

MAX. FRAME RATE

Full frame 4.2M

2048 x 2048

N/A

Full frame 2.2M

2048 x 1088

333 FPS

2K

2048 x 1080

336 FPS

HD 1080

1920 x 1080

336 FPS

SXGA

1280 x 1024

355 FPS

XGA

1024 x 768

471 FPS

HD 720

1280 x 720

502 FPS

SVGA

800 x 600

600 FPS

PAL

768 x 576

625 FPS

WVGA

752 x 480

748 FPS

VGA

640 x 480

748 FPS

QVGA

320 x 240

1460 FPS

STANDARD RESOLUTIONS

         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™         DEWESoft™

        measurement innovation        measurement innovation        measurement innovation        measurement innovation        measurement innovation        measurement innovation        measurement innovation 

Содержание DS-CAM-600

Страница 1: ...esoft com DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft measurement innovation measurement innovation measurement innovation measurement innovation measurement innovation measurement innovation measurement innovation ...

Страница 2: ... 6 2 3Resolutions Framerates 7 3Installation 8 3 1Installation in DEWESoft X 8 3 2Installation in DEWESoft 7 1 9 3 2 1Required files 9 3 2 2Step by step instruction 9 4FAQ 14 5Documentation Version 17 5 1Reference list 17 www dewesoft com Page I DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft measurement innovation measurement innovation ...

Страница 3: ...1000Base T connectible to every standard GigE Ethernet port and cable lengths up to 100m Dewesoft uses Smartek GigE SDK for communication with cameras that support GigE Vision standard IMPORTANT For best performance we recommend using a Core i5 CPU or better 4 GB RAM and SSD for storing data 1 1 System Requirements 1 Gigabit Ethernet port Good PC performance Core i5 CPU or better recommended 4 GB ...

Страница 4: ...or CMOSIS CMV2000 2E5M1PP Sensor type CMOS Sensor resolution H x W 2048 x 1088 Optical size Diagonal 12 7 mm 2 3 Pixel size in μm 5 5 x 5 5 Pixel data format JPEG Dynamic range 60 dB Shutter Electronic Global Shutter Shutter time 210 ns 90 s Exposure Linear 3Slope High Dynamic Range Scanning system Progressive Max frame rate at 640x480 specified for 600 in fact 748 are possible ADC bit depth 10 bi...

Страница 5: ...J45 Sync 4 pin LEMO EGG 00 304 CLL Conformity CE EN55022 class A EN61000 4 2 EN61000 4 3 EN61000 4 4 EN61000 4 6 FCC Part 15 class A RoHS GigE Vision 1 2 Illustration 1 Mechanical Drawing Page 3 17 www dewesoft com Document Version 1 0 0 DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft measurement innovation measurement innovation measurem...

Страница 6: ...Illustration 2 Connectors Illustration 3 IO connector pinout Note that pins 3 and 4 are used for clocking the camera Pins 1 and 2 are reserved for synching multiple DEWESoft instruments Document Version 1 0 0 www dewesoft com Page 4 17 DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft measurement innovation measurement innovation measuremen...

Страница 7: ...ta acquired by DEWESoftTM Note Even if the camera is connected synching to other SIRIUS devices is still possible because there are different pins used on the Synch Connector for triggering synching see also IO connector pinout Illustration 4 SIRIUS S Box with one SIRIUS on top connected to the DS CAM 600 PoE injector in between Page 5 17 www dewesoft com Document Version 1 0 0 LAN in Trigger cabl...

Страница 8: ...rument is equipped with a 37pin DSUB connector please see DEWETRON manual for pinout here for example is the pinout for the Orion 1622 100 see Reference List b Connect EXT_CLK1 pin 17 to CLK of the Lemo 00 connector of the camera pin 3 see 2 2 1 Connect DGND pin 37 to GND of the Lemo 00 connector pin 4 see 2 2 1 Set the pin EXT_CLK1 to Video in DEWESoft hardware setup Illustration 5 DEWETRON Orion...

Страница 9: ... 1088 Resolution Active pixels MAX FRAMERATE Full frame 4 2M 2048 x2048 N A Full frame 2 2M 2048 x1088 333 FPS 2K 2048 x1080 336 FPS HD 1080 1920 x1080 336 FPS SXGA 1280 x1024 355 FPS XGA 1024 x768 471 FPS HD 720 1280 x720 502 FPS SVGA 800 x600 600 FPS PAL 768 x576 625 FPS WVGA 752 x480 748 FPS VGA 640 x480 748 FPS QVGA 320 x240 1460 FPS STANDARD RESOLUTIONS DEWESoft DEWESoft DEWESoft DEWESoft DEW...

Страница 10: ...fullinstaller again and reinstall the option The last DEWESoft X fullinstaller can be downloaded here http www dewesoft com dewesoft7 developer downloads Illustration 7 DEWESoft X installing GigE camera driver Document Version 1 0 0 www dewesoft com Page 8 17 DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft measurement innovation measureme...

Страница 11: ...rtek_GigEVisionSDK Note that you must be administrator not just user with admin rights After installation be sure that Smartek GigE Vision Filter Driver is installed picture 1 under Local area Connection Properties HINT Note that Filter Driver filters out all packets that are not GigE on hardware level so the camera will work much faster than without the filter Page 9 17 www dewesoft com Document ...

Страница 12: ... camera is working you can run Smartek GigEVisionClient By right click on the camera you can also set IP to fix if you prefer Illustration 9 OK cameras can be used close the GigEVisionClient and start DEWESoft Document Version 1 0 0 www dewesoft com Page 10 17 DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft measurement innovation measurem...

Страница 13: ...lone or Master Illustration 10 Dewesoft Settings Hardware setup Analog Set device to Master or Standalone 7 Go to the Video tab and check GigE Vision and Use trigger Illustration 11 DEWESoft Hardware setup Video finds GigE camera s Page 11 17 www dewesoft com Document Version 1 0 0 Disable online compression Set compression after measurement DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DE...

Страница 14: ...nel Illustration 13 DEWESoft Channel setup Video HINT Ensure the firmware version of SIRIUS hardware is at least 4 5 0 16 otherwise please update the firmware This is done with the USB Firmware Upgrade Tool from http www dewesoft com download section 1 Also the Firmware File can be downloaded there Close DEWESoft start the Upgrade Tool select DEWESoft USB interface and firmware file and Start Afte...

Страница 15: ...with low dynamic range e g only light grey and dark grey shapes To improve the performance if using the camera on an Ethernet switch use Optimize for switched network Use the Advanced Setup if you want to change specific parameters of the camera such as the custom resolution and AOI area of interest for instance HINT If you change the frame rate after typing the value the input field gets a yellow...

Страница 16: ... device is set to Master or Standalone Disable Trigger in hardware setup and check if it s working in free run mode cable problem Camera not found not in GigEVisionClient not in DEWESoft Check if status LED on camera is green booting takes ABOUT 1 MINUTE If using a GigE Switch with PoE ports ensure output power is sufficient check status LED during operation Check Windows Firewall settings Camera ...

Страница 17: ...nd C Program Files Smartek GigEVisionSDK GenICam bin Win64_x64 to Windows system32 respectively Windows SysWOW64 Then start the GigEVisionClient from C Program Files Smartek GigEVisionSDK bin Win64 Cameras not working in DEWESoft If you followed the step by step installation procedure but still the camera is not working in DEWESoft you can try to copy the used dll s manually Copy the 32 bit dll s ...

Страница 18: ...d to enable Jumbo frames on your PC network card Jumbo frames are Ethernet packets larger than 1500 bytes This way less CPU time is spent for data reception therefore increasing performance and minimizing data loss Control panel network and internet view network status and tasks network and sharing center change adapter settings right mouse click on LAN connection Properties Configure Advanced Jum...

Страница 19: ... b Dewetron Orion AD board manual http download dewetron com dewe orion 1622 10x_106ex pdf _______________ For support please contact support at dewesoft com Page 17 17 www dewesoft com Document Version 1 0 0 DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft DEWESoft measurement innovation measurement innovation measurement innovation measurement in...

Отзывы: