Trouble-shooter’s Guide to Detect and
Eliminate Faults/Errors

The following table has been developed to help you detect and eliminate common faults
and errors. If you come across a fault/error that is not listed here, please contact your
supplier. All of these faults/errors may only be inspected and rectified by specialists
holding the qualifications as described on page 3.

19

Handset or
drive system
without function

- Handset or drive system

defective

- No supply voltage

- Contact your supplier/dealer

- Connect to the mains

Drives suddenly no
longer respond, no
movement takes place

- Thermoswitch on transformer or

in the control unit has been triggered

- Temperature fuse in transformer

defective

- Miniature fuse has been triggered

- Feeder cable (mains and/or handset/

slave drives) interrupted

- Leave the drive system in

rest position for approx.
20 - 30 minutes

- Contact your supplier/dealer

- Check the feeder cable, if

necessary restoring contact

Permanent or longer-
sounding warning signal
(CARE/HOSP only)

- Control unit is defective

- Contact your supplier/dealer

Function LED in the
Handset does not light
up when keys pressed
or glows constantly
(CARE L/CARE/HOSP)

- Fault in the control system
- End-of-travel position reached

- Blocking switch on locking device/

Supervisor or IPROXX SE/SE+ set at
“blocked"

- Drive load exceeded
- Battery backup system has reached

total discharge

®

- Contact your supplier/dealer
- Operate drive to move from

the end-of-travel position

- Release the blocking switch

- Relieve the drives
- Connect the system to the

mains to charge the batteries

IPROXX SE/SE+:
Mode LED lights up
in the blocked state

Mode LED does not
light up in the
released state

®

- Fault in the handset

No movements can
be carried out, by
using a handset with
blocking function

- Use the magnet to release

the handset

- Handset is blocked. (After a power

Failure, the IPROXX SE/SE+

handset automatically switches to
the blocked state)

- Locking device or Supervisor is

blocked (if being used).

®

Problem

Remedy

Possible Cause

- Contact your supplier/dealer

Содержание MCL

Страница 1: ...Issue 06 2005 Installation Instructions for the Manufacturer of the End Product Drive Control Unit MCL CARE L CARE HOSP...

Страница 2: ...Instructions the drive control units are described under the code designation In doing so you can prevent mistakes being made in installing or connecting the system which could result in to the drive...

Страница 3: ...rmed by a This being the case you should never carry out this work unless you are a or Conformity according to EC Directives The drive control system is supplied ex factory as a machine in accordance...

Страница 4: ...of the mains connection lead should be carried out at short term intervals and in particular each time it has been subjected to a mechanical load If the mains connection lead of the drive control uni...

Страница 5: ...switch Blocking devices Battery AG7 Functions CARE L CARE HOSP MEGAMAT MEGAMAT MEGAMAT MULTIMAT B23 MULTIMAT B23 MULTIMAT B23 MEGAMAT P MEGAMAT P MEGAMAT P MEGAMAT 2 MEGAMAT 2 MEGAMAT 2 MEDIX CARE ME...

Страница 6: ...re Length x width x height of the control system 175 x 139 x 90 mm Weight approx 2 2 kg Room temperature from 10 to 40 C Rel humidity from 30 to 75 Barometric pressure From 700 hPa to 1060 hPa 1 2 Dim...

Страница 7: ...battery backup system can also be connected optional If ordered the optional battery backup system is supplied ex factory as an individual unit i e not connected The connection diagram for the batter...

Страница 8: ...en the bed is being First connect the as shown see page 18 Make sure that any sockets not being used are sealed off with If connecting an e g for emergency adjustment fasten this to the application us...

Страница 9: ...ication plug s pull out safeguard 1 earthing connection fitting option 1 fitting option 2 Caution For your own safety For battery powered drive systems it is not sufficient to simply pull out the main...

Страница 10: ...ol element indicates that a button is pressed With the CARE HOSP versions it signals in addition Version CARE HOSP 10 If you have acquired the drive control unit with a please note following Charge th...

Страница 11: ...lways lights up when power is being supplied to the unit mains or battery backup system if included The display serves in addition as an indicator for the charge status of the battery backup system if...

Страница 12: ...drive should be immobilized Note keep the handset away from and 1mT The integral blocking device can be activated or deactivated Further information on the function of the blocking device can be found...

Страница 13: ...eatures the same pictographs as on the handset i e the depicted function is blocked by the blocking switch Blocking is symbolized by a closed shackle lock Furthermore the can include adjustment button...

Страница 14: ...pervisor allows the bed to be adjusted as well as the adjustment functions to be blocked The Supervisor has been designed as a two hand control thereby making it impossible to actuate adjustment inadv...

Страница 15: ...doing so check whether the function LED on the handset lights up and listen out for the acoustic signal Check the function of the blocking device as described on pages 12 to 14 to ensure its readines...

Страница 16: ...onding programs offered by DEWERT After correctly changing the mains connection lead check for safety and proper function Service shaft The service shaft allows you to easily replace damaged mains cab...

Страница 17: ...ive may only be opened by specialist personnel holding the qualifications as described on page 3 17 Graphical Symbols Conformity marks Protection classification II Application part of Type B Protectio...

Страница 18: ...unit e g AG7 DEWERT foot switch e g FP2 FP4 Pos Part designation Description 1 2 3 6 7 1 1 or or or This can otherwise result in damage to the drive control unit DEWERT bridge plug bridging the block...

Страница 19: ...le if necessary restoring contact Permanent or longer sounding warning signal CARE HOSP only Control unit is defective Contact your supplier dealer Function LED in the Handset does not light up when k...

Страница 20: ...onents cables metal plastic etc This system should be disposed of in accordance with the environmental regulations applicable in the respective country Information on this subject can also be obtained...

Страница 21: ...is machine may not be put into operation until conformity with the above Directive of the entire machine into which it is to be installed has been declared Partially applied harmonized standards of Ma...

Страница 22: ...d in the Installation Instructions as well as the designated use in other words which change the drive system in a considerable way make this Declaration of Conformity null and void Management DEWERT...

Страница 23: ...ive temperatures EN60601 1 main section 9 Operation outside the requirements and faults cases EN60601 1 main section 10 Constructional requirements EN60601 1 section 56 8 Power supply indicator option...

Страница 24: ...ing device supervisor or IPROXX SE SE EN60601 2 38 section 22 4 101 Control unit with keypad EN60601 2 38 section 36 Electromagnetic compatibility EN60601 2 38 section 52 4 Unintentional movement EN60...

Страница 25: ...25 Notes...

Страница 26: ...MCL 26 Notes...

Страница 27: ...27 Notes...

Страница 28: ...DEWERT Antriebs und Systemtechnik Weststra e 1 32278 Kirchlengern Germany Tel 49 0 5223 979 0 Fax 49 0 5223 75182 http www dewert de Info dewert de GmbH ID no 49343...

Отзывы: