background image

Dewert User Manual 6000E and 6800E – Updated 05.2021/1 MDR                                                                Page 7 of 15 

Lowerable side guard, for side rail (optional) 

This side guard can be placed and fixed on any side rail using a clamp. By loosening the toggle screw of the 
clamp, the position and height of the inserted side guard can be shifted and adjusted respectively. 
Complete lowering below the upholstery level can only be achieved if the guide points in the direction of the 
upholstery when the side guard is inserted into the clamp.    

Please note:

 Never use the side guard as a push handle. It has not been designed for this purpose. 

     
 

2.7 Special features of model 6100E   

The telescopic column of the table is attached directly underneath the foot end of the table. 
The upholstery support is a thick tubular steel frame which enables use without crossbars. This creates a free 
area between the end of the frame (end of lying part/beginning of head part) and the telescopic column.  
Here, a C-arm examination can be performed vertically in the chest to pelvis area of the lying patient. 
Optionally, the table can be equipped with an HPL support plate and removable upholstery. The upholstery is 
attached to the frame using hook-and-loop tape on the sides. To do this, affix the tape on the underside of the 
frame. The upholstery can be removed fully.  

 
 

3. Additional accessories (for user-specific table configuration, model dependent) 

 

 

 

Individually lockable castors Ø 100 mm   

 

Twin castors Ø 100 mm, centrally lockable  

 

Wheel-lifting/fixing mechanism  

 

Paper roll holder  

 

Lateral armrest  

 

Additional foot switch for low-voltage control  

 

Additional hand switch for low-voltage control  

 

DIN rails (stainless steel, 350 mm in length)  

 

DIN rail ac

cessories, e.g. fastening clamps, side guards, infusion poles… 

 

and more 

 

 

4. Technical data 

4.1 Individual models  

Models

 

6000E 

6050E 

6070E 

6075E 

 

6001E 

6051E 

6870E 

6875E 

 

6805E 

6810E 

 

 

Max. length (mm) 

1950 

1950 

2040 

2040 

Width (mm) 

700/800 

700/800 

700 

700 

Min. 

– max. height (mm) 

560 - 960 

560 - 960 

560 - 960 

560 - 960 

Adjustment time (motor) (s)  

24 

24 

24 

24 

Adjustment time, inclined 
position (motor) (s) 

20 

20 

Weight (approx., depending on 
equipment) kg 

80/85 

100 

110 

120 

Head part adjustment range  

+45° / -35° 

+70° / -25 

+35° / -30° 

+35° / -30° 

3-sectioned head part adjustment 
range (option) 

+45° / -35° 

+35° / -20° 

+35° / -20° 

Back part adjustment range 

+70° / -15° 

+70° / -15° 

Leg part adjustment range 

+50° / -55° 

+50° / -55° 

Adjustment range, inclined 
position 

0° / +25° 

0° / +12° 

Max. patient weight (kg) 

250 

250 

220* 

220* 

Содержание 6000E Series

Страница 1: ...ntended purpose 2 1 5 Information on setup and use 3 1 6 Commissioning 3 1 7 Safety instructions 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number 4 2 Operating manual 4 2...

Страница 2: ...th the contents of these instructions for use before using the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authoris...

Страница 3: ...y Remove the power cable from the film packaging on the underframe of the table and connect it properly to a permanently installed mains socket When routing the power cable make sure the cable cannot...

Страница 4: ...attached to the underframe on one long side of the table It provides information about key table data The following symbols are listed there by way of example their meanings are Read the operating ma...

Страница 5: ...as spring the head part must not be used for sitting Head part adjustment upwards and downwards positive and negative direction using a gas spring The head part is adjusted using a gas spring To opera...

Страница 6: ...l lifting fixing mechanism is not controlled centrally a foot lever must be operated on each short side of the table in order to lower the table onto its feet or to place it on the castors This option...

Страница 7: ...ort plate and removable upholstery The upholstery is attached to the frame using hook and loop tape on the sides To do this affix the tape on the underside of the frame The upholstery can be removed f...

Страница 8: ...te rest period There is therefore no risk of overloading the motor The mechanics of the tables have been designed with extensive safety margins in mind The max patient weight is provided with a 4 fold...

Страница 9: ...pid exchange of air is necessary To ensure this there are individual vent points on the underside of the upholstery which are hygienically sealed with special air compensation caps that assume a valve...

Страница 10: ...on In the event of any abnormalities take the table out of service immediately and inform the operator Replace damaged or worn components immediately and do not use the table until it has been repaire...

Страница 11: ...d to the manufacturer K H Dewert GmbH and to the competent authority of the Member State in which the user and or the patient resides Member State Competent authority Web Estonia Health Board Tervisea...

Страница 12: ...s later on the no longer used product must be rendered unusable immediately e g by disconnecting the power cable Please observe the local regulations for waste disposal and preferably recycle the mate...

Страница 13: ...fessionally trained persons who have been instructed in its use or who through experience with other similar medical devices have knowledge of its proper use taking into account possible hazards The d...

Страница 14: ...All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off to ensu...

Страница 15: ...ntral locking system also the grub screws of switch levers Safe braking effect Check with locked brakes by pulling and pushing the table Structural inspection Check screw connections Retighten all scr...

Отзывы: