Dewert User Manual 6000E and 6800E – Updated 05.2021/1 MDR                                                                Page 3 of 15 

performed by professionally trained persons who have been instructed in its use or who, through experience 
with other similar medical devices, have knowledge of its proper use, taking into account possible hazards. 
The device must only be moved within the room for cleaning or patient access. This device has been 
developed exclusively for use indoors and in normal ambient conditions and can be used in the following 
areas:  laboratories, medical practices, examination and treatment rooms, hospitals, clinics, physiotherapy 
practices, occupational therapy centres and doctors’ surgeries.  
This table is not classed as surgical furniture and must, therefore, not be used for surgical purposes. 
The expected service life is 10 years or 100,000 drive cycles (double stroke = 1x up and down). 

 
 

1.5 Information on setup and use

 

 

When packed, the device may be exposed to the following environmental conditions for approx. 3 months:   
     
Transport and storage temperature: -20° to + 50° 
Operating temperature: +10° to +40° 
Relative humidity: 30% to 75% 
Air pressure: 800hPa to 1060hPa 
       
When transporting the device in a vehicle, it must be secured properly against moving. To do so, lock the 
castors (optional equipment) and ensure further safety measures. 
 
When setting up the device, do not lift it at the head part, as this may damage the head part and/or the 
release mechanism. Hold and lift the table on the left and right-hand side of the frame. The table must stand 
securely on its feet or castors on a level, flat and solid surface. Before use, activate the brakes on the castors 
and make sure they are working properly.  
 

When transporting the table, take hold of the underframe, NOT the upper frame. 

 
 

1.6 Commissioning 

The device is ready for use upon delivery. 
Remove the power cable from the film packaging on the underframe of the table and connect it properly to a 
permanently installed mains socket. When routing the power cable, make sure the cable cannot be crushed, 
rolled over or otherwise damaged.  

 

Lock the castors (optional) or the wheel system (optional). As the operator, carry out a thorough and precise 
function check once the device has been set up. Prior to commissioning, clean the device and remove any 
contamination from transport. Make sure that no connecting cables from the hand or foot switch to the motor 
are trapped in the mechanism and thus damaged. 
Operation in potentially explosive atmospheres is not permitted.

 

 
 
1.7 Safety instructions  

This table may only be used for its intended purpose. Any other use is strictly prohibited and possibly 
dangerous. The manufacturer cannot be held responsible for damage caused by improper use. Patients may 
only be positioned in preparation for treatment/examination by professionally trained persons. 

Please note:

 This table is not classed as surgical furniture. 

 
Prior to and when adjusting the height of the table, make sure that no persons or objects are located in the 
adjustment range of the table and that nobody has their hands on the underframe

.

 

 
The following basically applies: Never reach into or under the frame of the table when adjusting the height. 
Height adjustment can result in injury if the user does not pay due care and attention. Therefore, take great 
care when performing this procedure. 

 

When adjusting the upholstery parts, make sure that no persons or objects are in the adjustment range. Make 
sure that no persons reach under the upholstery part or lean on the underframe. 
Important for the user: When adjusting the upholstery parts, do not reach under the spacers located beneath 
the upholstery parts. 

 

Always use both hands when adjusting the lying surface elements: Use one hand to operate the adjustment 
mechanism and the other hand for the lying surface adjustment. 

Содержание 6000E Series

Страница 1: ...ntended purpose 2 1 5 Information on setup and use 3 1 6 Commissioning 3 1 7 Safety instructions 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number 4 2 Operating manual 4 2...

Страница 2: ...th the contents of these instructions for use before using the device We recommend that you keep these instructions for use near the device for future reference The device may only be used by authoris...

Страница 3: ...y Remove the power cable from the film packaging on the underframe of the table and connect it properly to a permanently installed mains socket When routing the power cable make sure the cable cannot...

Страница 4: ...attached to the underframe on one long side of the table It provides information about key table data The following symbols are listed there by way of example their meanings are Read the operating ma...

Страница 5: ...as spring the head part must not be used for sitting Head part adjustment upwards and downwards positive and negative direction using a gas spring The head part is adjusted using a gas spring To opera...

Страница 6: ...l lifting fixing mechanism is not controlled centrally a foot lever must be operated on each short side of the table in order to lower the table onto its feet or to place it on the castors This option...

Страница 7: ...ort plate and removable upholstery The upholstery is attached to the frame using hook and loop tape on the sides To do this affix the tape on the underside of the frame The upholstery can be removed f...

Страница 8: ...te rest period There is therefore no risk of overloading the motor The mechanics of the tables have been designed with extensive safety margins in mind The max patient weight is provided with a 4 fold...

Страница 9: ...pid exchange of air is necessary To ensure this there are individual vent points on the underside of the upholstery which are hygienically sealed with special air compensation caps that assume a valve...

Страница 10: ...on In the event of any abnormalities take the table out of service immediately and inform the operator Replace damaged or worn components immediately and do not use the table until it has been repaire...

Страница 11: ...d to the manufacturer K H Dewert GmbH and to the competent authority of the Member State in which the user and or the patient resides Member State Competent authority Web Estonia Health Board Tervisea...

Страница 12: ...s later on the no longer used product must be rendered unusable immediately e g by disconnecting the power cable Please observe the local regulations for waste disposal and preferably recycle the mate...

Страница 13: ...fessionally trained persons who have been instructed in its use or who through experience with other similar medical devices have knowledge of its proper use taking into account possible hazards The d...

Страница 14: ...All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off to ensu...

Страница 15: ...ntral locking system also the grub screws of switch levers Safe braking effect Check with locked brakes by pulling and pushing the table Structural inspection Check screw connections Retighten all scr...