Solace Dialysis Scale Owner’s
39
Performing Load Cell Trim, Cont.
SETTINGS
PAGE 5
Calibrate Scale
Load Cell Trim
Power up Zero: No
Auto Locking: 4 sec.
<<<
Exit
>>>
8. Select
Load Cell Trim
from the Settings Page 5 screen.
LOAD CELL TRIM
1: 2.941 mV - +
2: 3.334 mV - +
3: 3.148 mV - +
4: 1.747 mV - +
0.00
Exit
9. Place a test weight of at least 10% of the scale capacity on each corner of the scale, one
at a time, and note the reading for each corner.
10. Place a test weight on the corner with the lowest reading and adjust the appropriate cell
mV reading until the reading agrees with the highest reading obtained in Step 9.
Press the
+
key to increase the mV reading
Press the
-
key to decrease the mV reading for each cell.