3300-0298-0M Rev E
y
Solace Dialysis Scale Owner’s
12
DIALYSIS OPERATION WITH PATIENT ID
Zero Weight Display
1.
Press the
key to turn the scale on.
2.
In Gross Weight mode (POUNDS/GROSS or KILOGRAMS/GROSS shown on weight
screen), if the scale is not showing zero weight on the weight screen press the
Ö
0
Õ
key.
3.
The weight screen will return to zero (0.0) and CENTER ZERO will be displayed to show
that the scale is ready to use.
Zero Weight Display with Item on Scale
1.
Press the
key to turn the scale on.
2.
Place the item (wheelchair, Dialysis chair, etc.) on the scale.
3.
The display will show the weight of the item on the scale.
4.
Press the
Ö
0
Õ
key.
5.
The weight screen will return to zero (0.0) and CENTER ZERO will be displayed to show
that the weight of the item has been "zeroed off" and the scale is ready to use.
Operation with a Patient ID
1.
Press the
key to turn the indicator on.
2.
If required, press the
Ö
0
Õ
key to zero weight screen (with or without item on the scale).
3.
Press the
key.
4.
The display will change to show the PATIENT ID screen.
P
PA
AT
TIENT IID
12345678
Key in the numeric
patient ID to be
transmitted when
the Print key is
pressed.
1
2
3
4
5
6
7
8
9
SAVE CANCEL
0
.
DEL
5.
Using the numeric keys, enter up to an 8-digit numeric identification number for the patient,
and then press the
Save
key.
6.
Assist the patient on the scale
7.
When weight is stable, a beep will sound, weight reading will automatically lock and
AUTO LOCK will be displayed. Note that the amount of time the reading will hold is
dependent upon the Auto Locking setting in Setup.
NOTE:
If more time is needed, press the
key to hold the locked weight reading.
ALWAYS assist the patient on and off the scale to ensure they do not fall.
NEVER leave a patient unattended while they are on the scale. Failure to
maintain control of the patient at all times can result in serious injury to the
patient and/or you.
Содержание Solace ID-3636S-855RMP
Страница 2: ...3300 0298 0M Rev E y Solace Setup and Calibration 2...
Страница 6: ......
Страница 8: ...3300 0298 0M Rev E y Solace Dialysis Scale Owner s 2...