3300-0298-0M Rev E
y
Solace Dialysis Scale Owner’s
30
1 = SmartCal, Cont.
8. The display will change to show “Working” for a few seconds and then change to show
“Set test weight over Cell 3”.
CALIBRATE SCALE
Set test weight
over Cell 3 and
press Save.
1
2
3
4
5
6
7
8
9
SAVE CANCEL
0
.
DEL
9. Place the calibrated test weight on the scale over Cell 3 and then press the
Save
key.
10. The display will change to show “Working” for a few seconds and then change to show
“Set test weight over Cell 4”.
CALIBRATE SCALE
Set test weight
over Cell 4 and
press Save.
1
2
3
4
5
6
7
8
9
SAVE CANCEL
0
.
DEL
11. Place the calibrated test weight on the scale over Cell 4 and then press the
Save
key.
12. The display will change to show “Working” for a few seconds and then return to
the Settings Page 5 screen.
13. Calibration is now complete.
14. Press the
Exit
key to exit Setup and return to the Weight screen.
Содержание Solace ID-3636S-855RMP
Страница 2: ...3300 0298 0M Rev E y Solace Setup and Calibration 2...
Страница 6: ......
Страница 8: ...3300 0298 0M Rev E y Solace Dialysis Scale Owner s 2...